Morgan MEDesign
, Inc. - Maintenance ~ Service Manual BASIC-ONE -
2005 Page 6
REPLACE AND SECURE KOM
POWER and HAND
CONTROLLER CABLE
IF PROBLEM PERSISTS-
CHECK AFFECTED MOTOR
CHECK AC MAINS
There may be weak AC voltages
present affecting the motor drives.
The STATUS lights will indicate the
voltage condition, refer to the
OPERATION MANUAL.
MONITOR CIRCUITS are O.K.
FILTER is O.K.
FAIRCHILD MONITOR is O.K.
LINE VOLTAGE is O.K.
+12 VOLT SUPPLY is O.K.
POSSIBLE OVERHEATED MOTOR
The motor drives have a duty cycle and when
this is exceeded will self-protect with a
thermal electrical breaker. WAIT 10
MINUTES and TRY OPERATION AGAIN
Possible KOM BOX failure or
HAND CONTROLLER
failure or HAND CONTROLLER
CABLE failure. Check for secure
connection.
WAIT 10 MINUTES MORE, if NO
OPERATION consider to REPLACE
the AFFECTED MOTOR
If MOTOR is O.K. consider
replacing KOM BOX or HAND
CONTROLLER