Warning: Means there is the possibility of personal injury to you or others.
Caution: Means there is the possibility of damage to the device.
Note: Provides information to further assist in the use of the device.
Warning(s)
• This manual gives important information about the use and safety of this device. Deviating from operating
procedures, misuse or misapplication of the device, or ignoring specifications and recommendations could
result in increased risk of harm to users, patients and bystanders, or damage to the device.
• Device captures and presents data reflecting a patient’s physiological condition that when reviewed by a trained
physician or clinician can be useful in determining a diagnosis; however, the data should not be used as a sole
means for determining a patient’s diagnosis.
• Users are expected to be licensed clinical professionals knowledgeable about medical procedures and patient
care, and adequately trained in the use of this device. Before attempting to use this device for clinical
applications, the operator must read and understand the contents of the user manual and other accompanying
documents. Inadequate knowledge or training could result in increased risk of harm to users, patients and
bystanders, or damage to the device. Contact Mortara Technical Service for additional training options.
●
To avoid the potential for spread of disease or infection, blood pressure cuffs should be cleaned after each
patient use.
• This device is intended for use with ambulatory patients. This device is not intended for use with unconscious
or sedated patients.
• Warning: During device configuration using the HDS (Hypertension Diagnostic Suite) Software if a “Top
Inflation Pressure” above 300mmHg is entered the system will automatically enter the value of 300mmHg.
Further, if a “Top Inflation Pressure” above 285mmHg is entered and sent to an AMBULO 2400 device the
HDS software returned an error message. If the AMBULO 2400 device is then disconnected from the system
running the HDS software, the device displays an error “SE 18” indicating “Data memory area is unexpectedly
erased” and the “Top Inflation Pressure” defaults to 180mmHg automatically.
• Use only approved blood pressure (BP) cuffs specifically intended for use with the Ambulo 2400.
• Do not fold, clamp, cut, or alter the pressure hose of the cuff or the monitor.
• Instruct the patients to check the limb connected to the cuff for adequate perfusion, circulation, and function.
Repeated NIBP measurements can lead to hematomas, limb ischemia, and other limb injuries. Kinked or
blocked hoses can lead to prolonged impairment of blood circulation and lead to injury.