EasyManua.ls Logo

MSE Soniprep 150 Plus - Page 21

MSE Soniprep 150 Plus
25 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OPERATING INSTRUCTIONS 21
USE THE CONTROL SWITCHES ON THE PROCESS TIMER ONLY WHILST
THE PROCESS TIMER IS PLUGGED INTO THE SONIPREP
APART FROM TURNING THE SONIPREPON AT THE MAINS AND SETTING
UP THE AMPLITUDE, THE PROCESS TIMER SHOULD BE USED TO
CONTROL THE SONIPREP ONLY.
( Please do not open the door whilst using the Process Timer because in
this mode the door interlock switch is not in circuit. )
Process Timer
THE WATTAGE METER (Part Number 46222-5020)
The wattage meter can be plugged into the back of the Soniprep to the 3 pin
din socket in order to see the power being dissipated into the sample.
The display on the meter shows the watts of power that the sample is receiving.
The display will show small up and down movements as the amplitude does
vary slightly depending on the peak or the trough of the sign wave.