L60iF Issue 8 July 2014 26
Calibration Menu
Warm Up
L60iF Calibration
when warm 018.0 C
Zero Flow
Zero flow reading
Please wait 9.9 sec
First Calibration Point
Cal Point 1 48 l/min
when flow set
Recording
Taking flow reading
Please wait 9.9 sec
Second Calibration Point
Cal Point 2 53 l/min
when flow set
Recording
Taking flow reading
Please wait 9.9 sec
Third Calibration Point
Cal Point 3 58 l/min
when flow set
Recording
Taking flow reading
Please wait 9.9 sec
Fourth Calibration Point
Cal Point 4 63 l/min
when flow set
Recording
Taking flow reading
Please wait 9.9 sec
Temperature Cal
Temperature Cal req? Cold flow rate 70.0 Hot flow rate 65.0
No Yes  adjust Save  adjust Save
Calibration data stored
Opening display
10:30 Fri 14/04/2000
last test run test