L60iF Issue 9 November 2018 3
Display resolution (Volume)
While the total volume sampled is less than 1000 litre, the display is in litres to one decimal
point, for example
738.3 litres
When the volume exceeds 1000 litres but is less than 100,000 litres, the decimal place is
dropped, for example
1429 litre or 83445 litres
When the volume sampled exceeds 100,000 litres, (100 m³) the display changes to cubic
metres to one decimal place, for example
0108.6 cu.m or 0743.9 cu.m
The maximum value is 9999.9m³. Therefore the sample will stop and the display shows OV
after approximately 115 days of continuous sampling.
The user has the option of displaying m³ at all times or using the Auto mode described above.
Calibration The unit can be calibrated using a certified flowmeter, a flow controller (restriction valve)
and the inbuilt calibration routine. A fuller temperature calibration can also be undertaken if
required.
Prime air Sliding vane pump (4 vanes)
mover
Motor type 1400 rpm induction; permanent capacitor.
& rating 220/240V ac 50Hz 0.65A; 110/120V ac 50Hz 1.3A
Motor Overload cutout trip; rating 2A. Typically cut out operates within
protection 40s if the motor stalls.
Dimensions 300 x 152 x 305 mm
Weight 14 kg
Accessories The following items are available:
Inlet adapters for Maypack or card-mounted filters
Extension hose to allow sampling head to be located up to 2.5m from the sampler.
Anglepoise trolley to support remote sampling head.
A variety of filter papers.
DC/AC Converter to enable the L60iF to operate from a 12V d.c. source in a vehicle. (Current
requirement is approximately 15A) This converter is only recommended for use in a vehicle.