EasyManua.ls Logo

Myron L ultrameter II 6PFC - User Manual

Default Icon
2 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
NOTE: Please read these instructions thoroughly and carefully — Damage caused by improper installation or mishandling will void
the warranties of the instrument and the new sensor.
CAUTION: Moisture trapped in instrument will cause damage. Careful adherence to the following procedure will signicantly reduce
the potential for failure due to moisture contamination.
Enclosed with this new sensor is a new orange O-ring, and a protective foam, install per instructions in III. 8. below. While the
instrument is open, we recommend you replace the battery with a 9V alkaline type. For clarity, this procedure is broken down into
four sections. Follow each section carefully to avoid damage. Refer to drawings on reverse as necessary.
I. Preparing and Opening Instrument Case.
1. Remove pH/ORP protective cap from sensor well, shake out uid and dry sensor well thoroughly.
2. If the instrument is wet, dry it thoroughly before proceeding, paying particular attention to the bottom screw
holes. Ensure hands are clean and dry. Moisture will damage the electronics.
3. Turn instrument over and set on a clean, dry surface.
4. Using a Phillips screwdriver, remove the 4 bottom screws.
5. Open instrument carefully, on Ultrameter it may be necessary to rock the bottom case side to side to release it
from the RS-232 connector. While removing bottom case ensure sealing gasket does not stick to bottom, but stays
in top case groove. If gasket comes out, handle carefully and reinstall in top case groove.
Exercise extreme care not to allow any water to enter the open case.
II. Disconnecting and Removing Old pH/ORP Sensor.
1. Locate the cable connectors and disconnect (unscrew using a 1/4 in. open-end wrench) both cables from the
circuit board noting position of RED pH cable. Red cable is on the left. See drawing B.
2. If you are right handed, hold the instrument in your left hand with the rounded end up and locate the pH/ORP
sensor and locking tab. See drawings B & C.
3. Using your left thumb, press the locking tab away from the sensor or toward the circuit board (the tab need
only be moved approximately 1/16 in./1,5 mm).
4. While holding locking tab back, pull straight out on the cables and body of the sensor, wiggling slightly if
necessary. Once the sensor moves upwards, you may let go of the tab and pull the sensor completely out.
5. Remove any pieces of the old seal and wipe out all dirty residue from the sensor well.
6. Thoroughly inspect and clean both top and bottom case (paying special attention to the case seal and O-rings),
screw wells and screws.
III. Preparing and Installing New pH/ORP Sensor
1. Install new ORANGE O-ring in sensor well. See drawing C.
Steps 2-4 should be performed AWAY from instrument.
2. Remove tape from around new sensor and protective boot. CAUTION: Protective boot is lled with a concentrated
saline storage solution that will cause damage if allowed to come in contact with internal components.
3. Carefully remove protective boot from new sensor by lightly squeezing top half of boot and executing a slow
twisting pull. Inspect to ensure 2 O-rings are on sensor body.
4. Rinse sensor with clean water and dry thoroughly. Inspect to ensure 2 O-rings are on sensor body.
5. Align sensor with case (it will only go in one way — see drawings B & C). Ensure both O-rings on sensor slide
evenly into sensor well. CAUTION: A damaged O-ring will allow water to leak into case.
6. Press sensor into sensor well until an audible snap is heard and locking tab is in place. See drawing C.
7. Reconnect the cables in their respective places: Red cable on the left, Black on the right. Tighten securely.
8. Remove existing protective foam (light colored) from the bottom case, peel backing from new protective foam, and
install inside bottom case in same location.
IV. Closing Instrument Case
1. Align cables so that they will NOT obstruct or become pinched while reinstalling the bottom case.
2. Ensure the sealing gasket is installed in the groove of the top case.
3. Replace bottom, aligning it with the groove.
4. Install and start the 4 bottom screws.
5. Tighten screws evenly and securely. A waterproof seal cannot be ensured unless screws are secure.
Uneven tightening will cause instrument to rock.
6. Turn instrument over and ll pH/ORP sensor well with Myron L pH/ORP Sensor Storage Solution. If unavailable,
use a pH 4 Buffer or a saturated solution of table salt and tap water.
7. Reinstall protective cap on pH/ORP sensor well.
Your instrument is now ready for calibration. See Operation Manual, Pocket Card or Bottom Label for calibration instructions.
2450 Impala Drive
Carlsbad, CA 92010-7226 USA
Tel: +1-760-438-2021
Fax: +1-800-869-7668 / +1-760-931-9189
www.myronl.com
RPRI r10-17
RPR pH/ORP SENSOR REPLACEMENT INSTRUCTIONS
For Ultrameter™ 6P, 3P, Ultrameter II 6P, 6Psi, 6Pfc, Ultrameter III 9P
PoolPro™ PS6, PS6si and D-6 Digital Dialysate Meter
All manuals and user guides at all-guides.com
all-guides.com

Questions and Answers

  • R
    Ryan MyersAug 4, 2025
    Why are my Conductivity/TDS/Resistivity readings unstable on Myron L Analytical Instruments?
    • T
      Thomas FranklinAug 5, 2025
      Unstable readings in Conductivity, TDS, or Resistivity measurements with Myron L Analytical Instruments can often be traced to dirty electrodes. Cleaning the electrodes can resolve this. Also, ensure that the test samples have appropriate conductivity, especially if they are greater than 1 megohm.
  • J
    Jesse NelsonAug 8, 2025
    What to do if Myron L ultrameter II 6PFC have no display even when the measurement key is pressed?
    • B
      Bridget FarrellAug 9, 2025
      If your Myron L 6P Analytical Instrument shows no display when you press the measurement key, the battery may be weak or not properly connected. Check the battery connections and replace the battery if necessary.
  • L
    Linda RichmondAug 13, 2025
    Why are the pH readings inaccurate on my Myron L 6P Analytical Instruments?
    • L
      Lisa YoungAug 13, 2025
      Inaccurate pH readings on your Myron L 6P Analytical Instruments can stem from several causes. It could be due to the need for pH calibration. Another potential cause is cross-contamination from residual pH buffers or samples in the sensor well, so ensure you clean the sensor well thoroughly. Also, using fresh pH buffers for calibration is important, as expired buffers can lead to inaccuracies.
  • J
    Jonathan MooreAug 17, 2025
    What to do if Myron L 6P Analytical Instruments show no response to pH changes?
    • A
      Angela AlvarezAug 17, 2025
      If your Myron L 6P Analytical Instruments show no response to pH changes, the sensor bulb might be cracked, or there could be an electromechanical short due to an internal crack. Inspect the sensor for damage and replace it if necessary.
  • S
    Samantha BarkerAug 21, 2025
    Why are resistivity readings much lower than expected on Myron L Analytical Instruments?
    • A
      Aaron MatthewsAug 21, 2025
      If the resistivity readings on your Myron L Analytical Instruments are much lower than expected, it could be due to contamination from a previous sample or from the pH sensor well. Cleaning the pH sensor well and electrodes may help. Also, carbon dioxide in the test sample can cause this issue, so ensure your sample is free from carbon dioxide.
  • M
    Matthew PierceAug 26, 2025
    What to do if Myron L 6P Analytical Instruments will not adjust down to pH 7?
    • S
      seth24Aug 26, 2025
      If your Myron L 6P Analytical Instruments will not adjust down to pH 7, it's possible that the pH/ORP sensor has lost KCl. Replenishing the KCl in the pH/ORP sensor should resolve this issue.
  • J
    Jamie DouglasAug 29, 2025
    Why do pH readings drift on Myron L 6P Analytical Instruments?
    • C
      Cynthia ChavezAug 30, 2025
      Drifting pH readings or slow response to changes in buffers/samples, or repeated display of “FAC” on your Myron L 6P Analytical Instruments, can be caused by several factors. It may be a temporary condition due to the sensor's “memory” of a solution in the pH sensor well for extended periods; in this case, allow the sensor to stabilize. Alternatively, the bulb might be dirty or dried out, so clean the pH sensor bulb. A clogged or coated reference junction can also cause these issues; clear any blockage in the reference junction.

Myron L ultrameter II 6PFC Specifications

General IconGeneral
Measurement ParameterspH, Conductivity, TDS, Salinity, Temperature, Resistivity, ORP, Free Chlorine
Salinity Range0 to 80 ppt
ORP Range±999 mV
Free Chlorine Range0.00 to 10.00 ppm
IP RatingIP67
pH Range0 to 14 pH
Conductivity Range0 to 200 mS/cm
TDS Range0 to 100 ppt
Temperature Range0 to 71 °C
Accuracy±1% of reading
Resolution0.01 pH, 1 mV (ORP), 0.01 ppm (Free Chlorine)
Calibration1-point calibration for ORP and Free Chlorine
DisplayLCD
Power Supply9V battery
Battery Life>100 hours
Operating Conditions0-55 °C (32-131 °F); 0-95% relative humidity, non-condensing
Storage Conditions-4 to 140 °F (-20 to 60 °C), 0 to 95% relative humidity (non-condensing)
Temperature CompensationAutomatic from 0 to 71 °C

Related product manuals