EPR User Manual N00020 L00034 1.0 03/2014 1
trophon
®
EPR
User Manual
ENGLISH

Table of Contents

Questions and Answers

  • K
    Karen PatrickSep 9, 2025
    What to do if the Nanosonics Trophon EPR chamber door is open and locked?
    • C
      Claudia HumphreySep 9, 2025
      If the chamber door of the Nanosonics Laboratory Equipment is open and locked, try selecting Restart on the LCD if the trophon is in sleep mode. Otherwise, power the trophon off and back on using the power switch. If this doesn't work, power off the trophon and follow section C4.3. Close the chamber door before powering the trophon back on.
  • A
    andrea67Aug 23, 2025
    Why is my Nanosonics Medical Equipment beeping?
    • K
      Kimberly GreeneAug 23, 2025
      If your Nanosonics Medical Equipment is beeping, follow the instructions on the screen. There might be a probe inside the device that needs to be removed.
  • D
    Destiny RodriguezAug 25, 2025
    What to do if the chamber door will not open on Nanosonics Medical Equipment?
    • K
      kelligreerAug 25, 2025
      If the chamber door on your Nanosonics Medical Equipment won't open, ensure the device has power. Follow any on-screen messages for instructions and confirm that any ongoing cycle (purge or disinfection) is complete. Also, check that the chamber door handle can be completely opened.
  • S
    Samantha WilcoxAug 26, 2025
    Why will the cycle not start on my Nanosonics Medical Equipment?
    • T
      Todd JohnsonAug 26, 2025
      If your Nanosonics Medical Equipment does not start the cycle, make sure the device has power and follow the instructions on the screen. Check that the probe is correctly loaded and the chamber door is closed. If you are at the ‘Clean and dry probe screen’ and the YES button will not register, or if the screen reads READY and the start button will not respond, contact your customer service representative.
  • D
    Dillon HensleyAug 28, 2025
    How to troubleshoot power issues with Nanosonics Medical Equipment?
    • C
      Christine SnyderAug 28, 2025
      If there is no power to your Nanosonics Medical Equipment, ensure that the device is plugged in and that you are using the correct power cable for your region. Also, make sure the power is switched ON at the wall outlet and that the power switch on the device is set to ON.
  • W
    William GreenAug 29, 2025
    What to do if the chamber door is open and locked on Nanosonics Medical Equipment?
    • C
      casekatieAug 29, 2025
      If the chamber door on your Nanosonics Medical Equipment is open and locked, try selecting RESTART on the LCD screen if the device is in sleep mode. Otherwise, power the device OFF and then back ON using the power switch. This should automatically unlock the chamber door. If these steps don't work, power OFF the device and manually override the door lock. Close the chamber door before powering the device back ON.
  • N
    Nina YoungSep 18, 2025
    What to do if the Nanosonics Laboratory Equipment is failing numerous cycles?
    • H
      Henry SheltonSep 18, 2025
      If the Nanosonics Laboratory Equipment is failing numerous cycles, record any error codes and the color of the chemical indicator, and then contact your customer service representative.
  • W
    wwilliamsSep 22, 2025
    How to troubleshoot no power to Nanosonics Trophon EPR trophon?
    • D
      Dr. Vincent GrantSep 22, 2025
      Ensure the Nanosonics Laboratory Equipment trophon is completely plugged in and switched ON at both the trophon power switch and at the wall outlet. Also, verify that the correct power cable for your region is being used.
  • D
    Dr. April HolmesSep 7, 2025
    How to fix a Nanosonics Trophon EPR Laboratory Equipment chamber door that won't open?
    • P
      Peter LeSep 8, 2025
      If the chamber door of your Nanosonics Laboratory Equipment will not open, ensure that the trophon has power and that there is no disinfection, warm-up, or purge cycle running.
  • D
    Dakota BrooksSep 4, 2025
    What to do if liquid is leaking from the Nanosonics Laboratory Equipment trophon?
    • B
      Brandon GoodSep 5, 2025
      If you see liquid or mist coming from the Nanosonics Laboratory Equipment trophon, do not come into contact with it. Wear appropriate PPE and ensure the area is well ventilated. Allow the trophon to complete the cycle, then turn it off and remove the power cord. Contact your customer service representative.

Summary

Part A - Introduction and Initial Setup

Section A1: Introduction to the trophon EPR

Provides general information about the trophon EPR, including validated probes and user training requirements.

Section A2: Important Warnings, Labels and Symbols

Details critical warnings, labels, and symbols associated with the trophon EPR for safe operation.

Section A3: Overview of Device Features

Describes the front, back, chamber, and side panels of the trophon EPR device, identifying key components.

Section A4: Installation Guide

Covers the essential steps for installing the trophon EPR, including positioning, cable management, powering on, and basic settings.

Part B - Routine Use and Maintenance

Section B1: Loading the Disinfectant

Explains how to handle and load the disinfectant cartridges, including safety precautions and spillage procedures.

Section B2: Routine High Level Disinfection Cycle

Details the process for performing a high-level disinfection cycle, including probe preparation and positioning.

Section B3: Purge Cycle

Outlines the reasons for and procedure to run a purge cycle to remove disinfectant and prepare the device for transport.

Section B4: Removing and Disposing of Used Disinfectant Cartridges

Provides instructions for safely removing and disposing of used disinfectant cartridges, including handling precautions.

Section B5: Incomplete or Failed Cycles

Addresses common issues like power failure or cycle faults, offering solutions and guidance for troubleshooting.

Section B6: Routine Care and Maintenance

Describes periodic cleaning procedures and when to contact for device service.

Section B7: Disposal of Device

Explains the process for decommissioning and disposing of the trophon EPR unit.

Part D - trophon Printer (Accessory)

Section D1: Introduction and Setup

Introduces the trophon Printer, its features, and initial setup procedures.

Section D2: Routine Care and Maintenance

Covers replacing the label roll and cleaning the trophon Printer.

Section D3: trophon Printer Troubleshooting

Provides guidance on connecting/disconnecting the printer and troubleshooting error messages.

Section D4: trophon Printer Disposal

Information on proper disposal of the trophon Printer for EU member states.

Section D5: Reorder Information

Details how to reorder trophon Printer supplies like the printer, cables, and label rolls.

Section D6: trophon Printer Technical Specifications

Lists the electrical, environmental, physical, and safety specifications for the trophon Printer.

Part E - Service and Warranty

Appendix 1: trophon EPR Technical Specifications

Provides detailed technical specifications for the trophon EPR, including electrical, environmental, physical, and EMC compliance.

Appendix 2: Product Warranty and Registration Form

Contains information on the product warranty terms, conditions, and the registration form.

Nanosonics Trophon EPR Specifications

General IconGeneral
BrandNanosonics
ModelTrophon EPR
CategoryLaboratory Equipment
LanguageEnglish