NI 6601/6602 Calibration Procedure 12 ni.com
6. Configure source of counter 1 to be PFI line 35.
7. Set the pulse delay for the single-pulse generation.
Traditional NI-DAQ Function Call LabVIEW Block Diagram
Call GPCTR_Change_Parameter with
the following parameters:
deviceNumber: The value set by MAX
gpctrNum:
ND_COUNTER_1
paramID: ND_SOURCE
paramValue: ND_PFI_35
Traditional NI-DAQ Function Call LabVIEW Block Diagram
Call GPCTR_Change_Parameter with
the following parameters:
deviceNumber: The value set by MAX
gpctrNum:
ND_COUNTER_1
paramID: ND_COUNT_1
paramValue: 2