NOTICE
This product may become more sensitive to
electromagnetic disturbances in the operational
environment when test leads are attached or
when the product is connected to a test object.
NOTICE
Operate this product only with shielded cables
and accessories. Do not use unshielded cables or
accessories unless they are installed in a shielded
enclosure with properly designed and shielded
input/output ports and connected to the product
using a shielded cable. If unshielded cables or
accessories are not properly installed and
shielded, the EMC specifications for the product
are no longer guaranteed.
NOTICE
The length of all I/O cables must be no longer
than 3 m (10 ).
NOTICE
You must install PXI EMC Filler Panels, NI part
number 778700-01, in all open chassis slots.
Electromagnetic Compatibility Standards
This product meets the requirements of the following EMC
standards for electrical equipment for measurement, control,
and laboratory use:
PXIe-5655 Safety, Environmental, and Regulatory Information
© National Instruments Corporation 9