Note Ensure that any connected external hardware is in an idle
state and not outputting signals during self-calibration. It is not
recommended to perform self-calibration while connected in direct
loopback configurations.
Note For the best image rejection and residual LO performance, run Self
Cal or Self Calibrate Range for the image rejection and residual LO portion
any time the module temperature changes beyond ±1 °C.
Performing PXIe-5842 Self-Calibration
You can self-calibrate your VST either interactively or programmatically; each
method yields the same self-calibration results.
Interactive methodsSelf-calibration in the InstrumentStudio application.
Programmatic methods—Self-calibration in a supporting programming
environment using an API directly.
NI-RFSA
Using the RFSA Self-Calibration programming example
Using self-calibration functions
NI-RFSG: using self-calibration functions
RFmx: using self-calibration functions
To get started, choose one of the self-calibration methods, and self-calibrate the
PXIe-5842.
Performing a VST Self-Calibration Using
InstrumentStudio
NI recommends you perform an initial self-calibration using InstrumentStudio.
© National Instruments
21
PXIe-5842 Getting Started

Other manuals for National Instruments PXIe-5655

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the National Instruments PXIe-5655 and is the answer not in the manual?

National Instruments PXIe-5655 Specifications

General IconGeneral
BrandNational Instruments
ModelPXIe-5655
CategoryTransceiver
LanguageEnglish

Related product manuals