3. Launch the Example Finder at Help » Find Examples... and navigate to the
example.
1. Select Hardware Input and Output » Modular Instruments » NI-
RFSA » Utilities » RFSA Self Calibration.
2. Open the example.
4. Complete the following steps to configure the example:
1. Select the device identifier assigned to the PXIe-5842 in MAX in the
[Resource Name] drop-down menu.
2. Set Clock Source to OnboardClock.
3. Set Self Calibration Step Operations to Perform All Self Calibration
Steps.
5. Run the VI.
PXIe-5842 stores the new calibration data in its nonvolatile memory if the self-
calibration is performed successfully.