Why is High O2 Alarm triggered on Newport Medical Instruments HT70?
- MMichael ThomasSep 18, 2025
A High O2 Alarm on your Newport Medical Instruments Fan can occur for several reasons: oxygen was increased prior to an intervention (e.g., suctioning) and not reduced, a low flow reservoir is in use and the patient’s minute volume has decreased or an airway/circuit leak has decreased during pressure control, NIV is on and Bias Flow was decreased, an oxygen concentrator is outputting a higher FIO2 than expected, the High O2 alarm is set inappropriately, or oxygen sensor calibration was not done appropriately. Adjust the oxygen setting back to the prescribed value and evaluate the patient, readjusting settings or the alarm as appropriate.

