Installation Guide
320907A-01 © Copyright 1995 National Instruments Corporation. All rights reserved. February 1995
SCB-100 100-Pin Shielded
Connector Block
This guide describes how to connect and use the SCB-100 100-pin shielded connector block with
100-pin data acquisition (DAQ) products.
Introduction
The SCB-100 100-pin shielded connector block is a shielded board with 100 screw terminals that
connect to the AT-MIO-64E-3, AT-MIO-16DE-10, or other products using the 0.050 series
shielded D type I/O connector.
The terminal block has 100 screw terminals for easy connection to signal wires. A cold-junction
compensation temperature sensor is included for use with thermocouples. When the SCB-100
100-pin shielded connector block is used with other 100-pin products, bypass the accessories (the
cold junction compensation temperature sensor and the signal accessory power LED) using the
six switches, leaving a generic 100-screw terminal connector block. The SCB-100 also has a
strain-relief bar for securing signal wires or cables.
What You Need to Get Started
SCB-100 100-pin shielded connector block kit
SCB-100 100-pin shielded connector block
SCB-100 quick reference label
SCB-100 100-Pin Shielded Connector Block Installation Guide
100-pin cable (SH100100)
100-pin board
Philips-head #1 and #2 screwdrivers
0.125 in. flathead screwdriver
Long-nose pliers
Wire cutters
Wire insulation strippers
Soldering iron and solder
Resistors and capacitors (specific to your application)
Question and Answer IconNeed help?

Do you have a question about the NI SCB-100 and is the answer not in the manual?

NI SCB-100 Specifications

General IconGeneral
BrandNI
ModelSCB-100
CategoryCables and connectors
LanguageEnglish

Summary

SCB-100 100-Pin Shielded Connector Block

Introduction

Overview of the SCB-100 shielded connector block, its features, and connectivity.

What You Need to Get Started

Lists the essential components and tools required for the initial setup of the SCB-100.

Switch Settings and Temperature Sensor Configuration

Disabled Temperature Sensor Switch Configuration

How to disable the cold-junction compensation temperature sensor using switches S4, S5, and S6.

Single-Ended Temperature Sensor Switch Configuration

Configuration for single-ended operation, connecting analog channel 0 to the temperature sensor.

Differential Temperature Sensor Switch Configuration

Configuration for differential operation, connecting analog channel 0 to the temperature sensor.

100 Generic Screw Terminals Switch Configuration

Bypassing circuitry for generic screw terminal use by switching S6 to the up position.

Temperature Sensor Output and Accuracy

Specifications

Signal Connection

Removing the Board

Application Hints

Soldering, Desoldering, and Cutting Jumpers

Guidance on modifying the SCB-100 board using soldering and jumper traces.

Connecting Nonreferenced or Floating Signal Sources

How to connect floating signal sources to the SCB-100 for accurate measurements.

Connecting Ground-Referenced Signal Sources

Guidance on connecting grounded signal sources to the SCB-100.

Using the SCB-100 Board for Thermocouple Measurements

Instructions and considerations for using the SCB-100 with thermocouples.

Optional Input Filtering and Broken Thermocouple Detection

Sources of Error