NICE RECOVERY SYSTEMS
GUIDANCE AND MANUFACTURER’S DECLARATION  ELECTROMAGNETIC IMMUNITY
The NICE1 is intended for use in the electromagnetic environment specied below. The customer or the user
of the NICE1 should assure that it is used in such an environment.
IMMUNITY TEST IEC 60601 TEST LEVEL COMPLIANCE
LEVEL
ELECTROMAGNETIC
ENVIRONMENT  GUIDANCE
Conducted RF IEC
61000-4-6
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 Vrms
3 V/m
Portable and mobile RF communi-
cations equipment should be used
no closer to any part of the Model
006, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1.2 √ P
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
where P is the maximum output
power rating of the transmitter in
watts (W) according to the trans-
mitter manufacturer and d is the
recommended separation distance in
meters (m).
Field strengths from xed RF
transmitters, as determined by an
electromagnetic site survey,a should
be less than the compliance level in
each frequency range.b
Interference may occur in the vicinity
of equipment marked with the fol-
lowing symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption
and reection from structures, objects, and people.
a Field strengths from xed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoret-
ically with accuracy. To assess the electromagnetic environment due to xed RF transmitters, an electromag-
netic site survey should be considered. If the measured eld strength in the location in which the Model 006
is used exceeds the applicable RF compliance level above, the Model 006 should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or
relocating the Model 006.
b Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than 3 V/m.