9
NICERECOVERY.COM
Level 2 = 54F (12.4C)
Level 3 = 50F (10C)
Level 4 = 46F (7.7C)
Level 5 = 42F (5.5C)
• Use Operating Temperature: 60°F - 80°F (15.5°C - 26.7°C)
• Relative Humidity : 10% - 95% non-condensing
• Storage/Transportation Temperature Range: 40°F - 105°F (4°C - 41°C)
• Maximum Sound Pressure Level: 75dBA (with air pump running)
• Maximum Altitude: 4000m (13,123 ft)
• Display accuracy: +/- 2%
2.3 MEDICAL SYSTEM SPECIFICATIONS
• Specified Medical Grade Power Supply. The following
Power Supply is specified for use with the NICE1 unit:
o Class II Power Supply: DC output rated 15Vdc 12A
NOTE: A detachable type power supply cord (rated 125V 10A) is
provided with each power supply, suitable for use in the USA and Canada.
• NICE1 Cooling Unit: Electrical Input Rating of 15Vdc 12A
• Patient Therapy Wrap (patient applied part): Classification
of Patient Applied Part, Type BF
• Insulated Air/Water Hose Assembly for connection between
Control Unit and Patient Therapy Wrap
• Leakage Currents (per IEC 60601-1):
o Touch (Enclosure) Current:
<28uA in Normal Condition
<50uA in Single Fault Condition
o Patient (Applied Part) Leakage Current:
<28uA in Normal Condition
<50uA in Single Fault Condition
2.4 OPTIONS
• Non-Sterile Patient Therapy Wraps (knee wrap, etc.)
• Carrying Case
Caution: Consult accompanying documents Lot or Batch Number
Refer to User Manual and Labels.
Follow instructions for use.
Serial Number
Caution: Federal law restricts this device to sale
by or on the order of a physician.
OFF position for the DC Power Switch
at the rear of the Control Unit
Type BF Patient Applied Part (therapy wrap) ON position for the DC Power Switch
at the rear of the Control Unit
SN
LOT
RxOnly
!