Chapter 8 Functions and Operations of the Devices
125
No. Operation Section Function
(1) Left foot pedal Left rear foot pedal (motorized focus unit Z-coordinate DOWN switch)
Drives the motorized focus unit downward.
When pressed, lowers the unit by 1.6 μm. Press and hold to move the unit continuously.
The speed changes based on the magnification and nosepiece address. (See the table
below.)
If a nosepiece address has not been set, the unit moves at the maximum speed.
Left front of the foot pedal (motorized focus unit Z-coordinate UP switch)
Drives the motorized focus unit upward.
When pressed, raises the unit by 1.6 μm. Press and hold to move the unit continuously.
The speed changes based on the magnification and nosepiece address. (See the table
below.)
If a nosepiece address has not been set, the unit moves at the maximum speed.
Coarse
Magnification
SMZ25 0.63 to 5.0 5.1 to 10.0 10.1 to 15.75
SMZ18 0.75 to 4.0 4.1 to 8.0 8.1 to 13.5
Nosepiece
address
1
10 (mm/second)
2
Fine
Magnification
SMZ25 0.63 to 5.0 5.1 to 10.0 10.1 to 15.75
SMZ18 0.75 to 4.0 4.1 to 8.0 8.1 to 13.5
Nosepiece
address
1 5 (mm/second) 4 (mm/second) 2 (mm/second)
2 2 (mm/second) 1 (mm/second) 0.5 (mm/second)
Extra fine
Magnification
SMZ25 0.63 to 5.0 5.1 to 10.0 10.1 to 15.75
SMZ18 0.75 to 4.0 4.1 to 8.0 8.1 to 13.5
Nosepiece
address
1 0.5 (mm/second) 0.4 (mm/second) 0.2 (mm/second)
2 0.2 (mm/second) 0.1 (mm/second) 0.0.5 (mm/second)
(2) Right foot pedal Right rear of the foot pedal (motorized magnification UP switch)
Increases the magnification by one step. The step size depends on the zoom step size
setting.
Press and hold to increase the magnification continuously.
The zoom step size depends on the setting of the zoom step switching button on the remote
controller.
Right front of the foot pedal (motorized magnification DOWN switch)
Decreases the magnification by one step. The step size depends on the zoom step size
setting.
Press and hold to decrease the magnification continuously.
The zoom step size depends on the setting of the zoom step switching button on the remote
controller.
(3) Foot rest Used as a foot rest.
(4) CTRL connector Attach to the FOOT connector of the P2-CTLA Control Box using the supplied cable.
(5) PHOTO connector Connect the AZ-PCR Photo Release. (See “10.3 AZ-PCR Photo Release” in this chapter.)