Chapter 3 Assembling
35
Filter and mirror attachment to the P2-EFLC Filter Cube
Filters of the following sizes and a mirror of the following size can be attached to the P2-EFLC Filter Cube:
Excitation filter (EX): External diameter 25 mm, Thickness 6 mm or less
Dichroic mirror (DM): Dimensions 36 mm x 25.5 mm, Thickness 1 mm or less
Barrier filter (BA): External diameter 25 mm, Thickness 4 mm or less
(1) Remove the top part and cover from the P2-EFLC.
1) Loosen and remove the two top part fixing screws from the top surface of the filter cube to take off the top
part.
2) Loosen and remove the four cover fixing screws from the right and left sides of the filter cube to take off
the cover.
Entire P2-EFLC Top part removed Cover removed (bottom part)
P2-EFLC disassembly
* See the section view on the right for the reference
plane of each part.
Reference plane
(2) Bond to attach the barrier filter to the bottom part.
1) Push the barrier filter into the position shown in
the figure of the bottom part until it hits the
bonding surface. Observe the manufacturer’s
instructions for the orientation of the barrier
filter.
Barrier filter setting - 1
2) Fill in the gap between the bottom part and the
barrier filter with adhesive.
Barrier filter bonding - 1
(3) Bond to attach the barrier filter to the top part.
1) Push the barrier filter into the position shown in
the figure of the top part until it hits the bonding
surface. Observe the manufacturer’s
instructions for the orientation of the barrier
filter.
Barrier filter setting - 2
Top part
Bottom
part
Cover
Top part fixing screws
Cover fixing screws
(also two screws on
the opposite side)
BA bonding surface
BA bonding
surface
DM bonding surface
Top par
Bottom par
DM abutting surface
EX bonding
surface
Bond along the
circumference.