Chapter 8 Preventive care
000191-16 NIOX VERO
®
User Manual English 31
2. Remove the used handle from the
instrument by pushing the socket into
the device and gently pull out the tube.
3. Discard the breathing handle.
CAUTION! The breathing handle contains
potassium permanganate and should be
disposed of as hazardous waste in
accordance with local waste disposal
regulations.
Do not re-use an expired breathing handle.
4. Attach a new breathing handle to the
instrument by pushing the tube into
the socket until the triangle is no
longer visible.
5. Select the Settings button in the main
menu.
6. Select Breathing handle button.
7. Select the Reset Breathing handle
button.
8. The breathing handle information view opens to confirm replacement of
the breathing handle. Select the OK button to confirm insertion of a new
breathing handle and to set the remaining measurements to 1000 and
expiration date one year from the current date.
Note: The Return button returns to settings view without
registering
change.
8.2.2 Exchange of NIOX VERO® Sensor
1. Turn off the instrument.
2. Open the compartment on the back of the instrument using a screwdriver.
Turn the swivel to release the sensor.
3. Remove the old sensor.
4. Replace with a new sensor.
WARNING! Make sure to not
touch or clean the white sensor
membrane.
WARNING! Be careful when
opening the sensor can. The
inside of the opening may have
sharp edges.
5. Turn the swivel to lock.
6. Replace the compartment lid.
CAUTION! Make sure there is no foreign material or particles in the sensor
compartment before closing it.
8.2.3 Change battery
If the rechargeable battery is no longer charging properly, malfunctioning, or
requires charging more frequently than normal, then it needs to be replaced.
Note: Only rechargeable batteries supplied by Circassia may be used.
(Type No BJ-G510039AA, Article No 12-1150)