EasyManua.ls Logo

Norma Icon-5 - Step 10: Verify Mixer operation; Mixer operation verification; Step 11: Reagent sensors and connections; Reagent sensor tube placement

Norma Icon-5
22 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
19
TDMG0002 Icon-5 Maintenance Guide Version: 02 Release date: 2020-04-27
Step 10: Verify Mixer operation
Frequency:
1 year or 10000 samples
part#: N/A
Get access to service level and check if the mixer is working
properly.
In case of any error, replace the mixer.
Step 11: Reagent sensors and connections
Frequency:
1 year or 10000 samples
part#: N/A
Make sure all tubes are properly placed in
the reagent sensors.
It should “click” in.
In case there is a liquid detection problem, clean the sensors with a brush. Run priming
sequence to see if the sensors are working properly.
Check reagent pickup tube connections to analyzer and reagent containers. In case a
connection is loose damaged cut 10mm from the end of tubing and reconnect.