Turn on the system, attach the foot-pedal, and place the sterilized EHL probe and extender cable onto the sterile field.
Connect the EHL probe to the extender cable, ensuring the square black ferrite aligns opposite the O.R. circulator connection.
The nurse connects the extender cable to the 'PROBE' port, displaying the probe size for precise energy delivery.
Insert the EHL probe into the endoscope working channel and align the probe tip with the stone for fragmentation.
Start with low power (suggested 30-40%) and increase as needed for effective stone fragmentation.
Adjusts shot delivery speed; higher frequency for stones that move, lower for stable stones. Highest is 30/sec.
Selects shots per foot-pedal depression; max 60 shots over 2 seconds. Suggested setting: 3-5 shots.
Use 'Standby' to prevent accidental firing; 'Total Shots' displays discharged shots.