The AUTOLITH System
Quick Reference and Information Guide
Getting STARTED
Turn on the power ON switch after plugging the AUTOLITH
®
System into a
proper power outlet. Attach the foot-pedal to the front panel foot-pedal port and
place the sterilized EHL Probe and Extender Cable onto the sterile field using
proper sterile technique.
Connecting the EHL Probe
Connect the EHL probe to the end of the extender cable opposite
the square black ferrite. Pass the remaining end of the extender
cable to the O.R. circulator.
EHL Probe Recognition
The O.R. circulating nurse should then connect the remaining end of the
extender cable to the front panel PROBE port. Notice the Connect
Probemessage change to indicate the size of the EHL probe that has
been connected, i.e. 1.9Fr. (.63mm). The AUTOLITH
®
System will now
deliver the precise amount of energy specific to the size of the
connected EHL probe.
The Endoscope and EHL Probe
The doctor should now prepare to fragment the stone. Careful insertion of the EHL probe into the
appropriate endoscope working channel is now accomplished and treatment of the stone is ready
to start. Using direct visualization, the EHL probe tip and the stone are sighted. Normal
physiological SALINE (premixed bags of .9% normal saline) should be used for irrigation at all
times.
Power
Start LOW and increase the power if needed for
effective stone fragmentation.
The total power range is 10% to 100%
Suggested to start: 30%-40% Power
®
Indicated for:
Urinary & Biliary
Stones
BiPolar Electro-Hydraulic Lithotripsy

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Nortech AUTOLITH IEHL and is the answer not in the manual?

Nortech AUTOLITH IEHL Specifications

General IconGeneral
BrandNortech
ModelAUTOLITH IEHL
CategoryMedical Equipment
LanguageEnglish

Related product manuals