EasyManua.ls Logo

North Atlantic VXI 65CS4 - User Manual

Default Icon
38 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Model 65CS4
MULTI-CHANNEL SYNCHRO/RESOLVER
STIMULUS AND MEASUREMENT
North Atlantic Industries, Inc. 631.567.1100/631.567.1823 (fax) 8/14/2013 65CS4_A001 Operations Manual Rev C6
110 Wilbur Place, Bohemia, NY 11716 www.naii.com
Cage Code: 0VGU1 Page 1 of 38
VXI bus
Four (4) Instrument Grade Synchro/Resolver Measurement Channels
Four (4) Instrument Grade Synchro/Resolver Stimulus Channels
or
Eight (8) Embedded Grade Synchro/Resolver Stimulus Channels
FEATURES
INSTRUMENT Grade (High Accuracy) and/or EMBEDDED Grade (Moderate Accuracy w/ High Channel Count)
Multiple functions on a single slot VXI card
±0.005° Instrument Grade Measurement and Stimulus Accuracy
±0.015° Embedded Grade Stimulus Accuracy (higher channel density / lower accuracy)
47 Hz to 4,000 Hz (see part number).
(for 47 Hz to 10 KHz or 20 KHz, contact factory )
User programmable output voltages
2.2 VA Stimulus Outputs
2.2 VA Programmable Reference Generators
Simultaneous and independent Measurement and Simulation
Single-Speed or Multi-Speed Programmable for Measurement and Simulation
Programmable Multi-Speed Ratios (2 to 255)
Dynamic address configuration
VXIbus data rate of 2 megabytes/sec
Data is processed within 100s
Self-Test capability
As direct replacement for 5410C-89, use P/N 65CS4-A2200G0-A2100G0-M and cable harness 07-0018.
As direct replacement for B5410C-89-1, use P/N 65CS4-A2200G0-A2101G0-M and cable harness 07-0021.

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the North Atlantic VXI 65CS4 and is the answer not in the manual?

Summary

FEATURES

DESCRIPTION

GENERAL ARCHITECTURE

SPECIFICATIONS

Motherboard - General

General specifications for the motherboard, including DSP, data transfer, and ESD protection.

Instrument Grade Measurement, Input Module(s)

Detailed specifications for Instrument Grade measurement capabilities, input modes, range, resolution, and accuracy.

Instrument Grade Stimulus, Output Module(s)

Specifications for Instrument Grade stimulus output, including channels, mode, resolution, accuracy, and drive.

Reference Generator

Specifications for the Reference Generator on the Instrument Grade Daughter Board, covering voltage, resolution, and accuracy.

Embedded Grade Stimulus, Output

Specifications for Embedded Grade stimulus output, including channels, mode, resolution, accuracy, and drive.

Overall Unit Power Requirements; Matrix

Power consumption details for motherboard and daughter boards, listing average current draw for +5V and +/-12V.

CARD ADDRESS

FRONT PANEL STATUS INDICATORS

PROGRAMMING

Native Syntax

Describes the Native Syntax for programming the 65CS4, including addressing and self-test procedures.

Software; Driver Support

Information on the VXI Software Support Kit (SSK), IVI driver, and LabWindows application provided.

Formal Syntax Notation

Defines the notation used in command syntax, including delimiters for mandatory/optional items and grouping.

General Command Definition

Explains the general structure and types of values (<value>, <scientific number>, <decimal number>, etc.) used in commands.

SIMULATOR COMMANDS

Angle

Commands for setting and querying the output angle in degrees for simulator channels.

DC Scale

Commands for setting and querying the DC output scale, defining the full-scale voltage and degrees/sec/V.

Rotation Rate

Commands to set and query the rotation rate in revolutions per second for simulator channels.

Rotation Stop Angle

Commands to set and query the stop angle for simulator rotation.

API COMMANDS

Angle

Commands to query the API angle in degrees for measurement channels.

DC Scale

Commands to set and query the DC output scale for API measurement channels.

Relay Function (API)

Commands to set and query the state of I/O isolation relays for API channels.

Measurement Mode

Commands to set and query the API channel update mode (LATCH/TRACK).

REFERENCE COMMANDS

Reference Generator Frequency

Commands to set and query the internal reference generator frequency in Hz.

UTILITY FUNCTIONS

Self-Test

Command to initiate self-test and query its status.

Reset

Command to reset the unit to its power-up default state.

SIMULATOR TRIGGER CONTROL DEFINITIONS

Explains trigger sources (BUS, INT, EXT, TTL) and slopes for simulator rotation control.

Instrument Setup Queries

Mate;CIIL Syntax

COMMANDS

Lists the available commands for Mate/CIIL syntax, such as FNC, CLS, OPN, RST, STA, CNF, IST, INX, FTH.

NOTATION

Defines the notation used in Mate/CIIL syntax, including delimiters for mandatory/optional items and grouping.

SENSOR FORMAT

STIMULUS FORMAT

FORMAT

Stimulus Reset

Details the RST command for stimulus functions, including results and cleared messages.

Sensor Reset

Details the RST command for sensor functions, including results and required MCHAR matching.

ERROR MESSAGES

EXAMPLE

Provides practical examples of using commands for setting synchro output, reading angles, and controlling rotation.

CALIBRATION

CALIBRATION VERIFICATION TEST SET-UP

Outlines the procedure for verifying unit performance, requiring external test equipment.

Internal Reference Checkout Procedure

Step-by-step guide for checking the internal reference functionality using specific commands and equipment.

Single-Speed Generation (Simulator) External Reference Checkout Procedure

Procedure to verify simulator performance using an external reference, including angle and API readings.

Generation (Simulator) Angle Rate Checkout

Procedure to verify the angle rate generation capabilities of the simulator.

Single-Speed Measurement (API) Internal Reference Checkout Procedure

Procedure for verifying API measurement using internal reference, covering Synchro and Resolver modes.

Single-Speed Measurement (API) External Reference Checkout Procedure

Procedure for verifying API measurement using an external reference, covering Synchro mode.

Two-Speed Measurement (API) Checkout Procedure

Procedure to verify API measurement for two-speed applications.

Generator (Simulator) DC Rate Output Checkout Procedure

Procedure to verify the DC rate output functionality of the simulator.

FIELD ALIGNMENT

Figures for Calibration Verification Test Set-up

CONNECTOR CONFIGURATION

Front Panel Connector Placements; Designations

PART NUMBER DESIGNATION

APPENDIX A - CE

INFORMATION FOR UNITS SOLD WITHIN THE EUROPEAN UNION

Information for units sold in the European Union, including classification, installation, safety, and maintenance.

DECLARATION OF CONFORMITY

REVISION HISTORY

North Atlantic VXI 65CS4 Specifications

General IconGeneral
Model65CS4
Form FactorVXI
CompatibilityVXIbus
CoolingForced Air
Operating Temperature0°C to 55°C
Humidity5% to 95% non-condensing
CategoryComputer Hardware
Storage Temperature-40°C to 70°C
Slot SizeC-size