4) Keep the components in the autoclave pouch until ready to use.
5) Store in a dry and dust-free location up to the period specified by the sterilization pouch manufacturer. If sterility cannot be
confirmed, sterilize again prior to use.
Steam Sterilization Cycle
Type Gravity Displacement Pre-vacuum
Minimum Temperature 121˚C 132˚C 135˚C 135˚C
Full Cycle Time 20 minutes 15 minutes 10 minutes 10 minutes
Minimum Dry Time 10 minutes 10 minutes 10 minutes 15 minutes
<When sterilizing components except for a Tip>
CAUTION
Do not exceed 138˚C during sterilization.
CAUTION
Do not sterilize by ultraviolet ray. The handpiece could discolor.
If autoclaved with other instruments stained with chemical solution, it could strip the plating
and make the surface black.
Do not autoclave any parts (the Control Unit, AC Adaptor, Foot Control, Handpiece Cord,
Tip Card, Tip Power Guide, O-Ring). Other than those that can be subjected to autoclave
sterilization. Perform alcohol disinfection to the Control Unit, AC Adaptor, Foot Control,
Handpiece Cord, Tip Card, Tip-Power Guide including after every patient.
Do not wipe with, or clean or immerse in, high acid water or sterilizing solutions.
Diamond coated Scaler Tip is for single use only. Do not sterilize and use again.
Keep the product in suitable atmospheric pressure, temperature, humidity, ventilation, and
sunlight. The air should be free from dust, salt and sulphur.
Do not touch the product immediately after autoclaving may cause injury.