Operation Manual
Pro O
2
Pro O
2
Remote
Oxygen Analyzers
01.11
Andersen Medical Gas
12 Place Lafifitte
Madisonville, LA 70447
http://www.TheMedicalGas.com
1-866-288-3783

Other manuals for Nuvair Pro O2

Questions and Answers

  • T
    Tracey SmithAug 3, 2025
    Why Nuvair Measuring Instruments reading is erratic?
    • J
      james31Aug 3, 2025
      An erratic reading on your Nuvair Measuring Instruments can occur due to several reasons. It might be caused by pressure on the sensor, so check the flow. Radio transmission interference could also be the culprit, in that case, move the unit away from potential sources of interference. Another reason may be that the sensor is old or faulty, consider changing the sensor. Finally, condensation on the sensor can cause this issue; try drying it in the air.
  • V
    Victor SmithAug 8, 2025
    What to do if Nuvair Measuring Instruments has no display?
    • T
      tcoxAug 8, 2025
      If your Nuvair Measuring Instruments' display is blank, here are a few things to check: First, ensure the device is switched on. If it is already on, inspect the display connection for any issues. Also, verify the battery connection is secure.
  • R
    Ricky PatrickAug 13, 2025
    What to do if Nuvair Measuring Instruments reading does not change when calibration knob is turned?
    • B
      Brian IbarraAug 13, 2025
      If the reading on your Nuvair Measuring Instruments isn't changing when you turn the calibration knob, examine the connections for any faults. It's also possible that the sensor has failed and needs replacing.
  • C
    Carl MartinezAug 18, 2025
    What to do if Nuvair Measuring Instruments shows battery symbol?
    • C
      Courtney NelsonAug 19, 2025
      If the battery symbol appears on your Nuvair Measuring Instruments, it indicates a low battery. Replace the battery.
  • L
    Linda Young MDAug 25, 2025
    Why Nuvair Measuring Instruments reading drifts?
    • J
      Jacob CarpenterAug 25, 2025
      If the reading on your Nuvair Measuring Instruments drifts, it might be due to a rapid temperature change. Stabilize the temperature and then recalibrate the instrument.
  • M
    Michael ThomasAug 28, 2025
    What to do if Nuvair Measuring Instruments shows zero reading?
    • M
      Melissa JonesAug 29, 2025
      If your Nuvair Measuring Instruments displays a zero reading, start by checking the sensor connection to ensure it's properly connected. If the connection is good, the sensor might have expired and require replacement.

Summary

1.0 Introduction

Signal Words and Safety Alerts

Explains DANGER, WARNING, CAUTION, and NOTICE signal words used in the manual.

2.0 System Description

Pro O2 vs. Pro O2 Remote Analyzers

Differentiates Pro O2™ and Pro O2 Remote™ models, their features, and applications.

Controls, Display, and Oxygen Sensor

Details the analyzer's physical controls, display interface, and the electrochemical oxygen sensor.

Battery and Flow Adapter Cap

Covers the battery specifications and the usage of the Flow Adapter Cap accessory.

Flow Diverter for Pro O2 Remote

Explains the installation and function of the Flow Diverter for the Pro O2 Remote™ model.

3.0 Calibration

Calibration Methods and Conditions

Discusses calibration procedures, environmental conditions, and correction values.

Calibrating the Pro O2 Analyzer

Provides step-by-step instructions for calibrating the Pro O2™ model in air.

Calibrating the Pro O2 Remote Analyzer

Provides step-by-step instructions for calibrating the Pro O2 Remote™ model in air.

4.0 Operation

Analyzing Gas with Pro O2

Details the process for analyzing gas samples using the Pro O2™ model.

Analyzing Gas Cylinder Samples

Illustrates the direct method for analyzing gas cylinder contents with the Pro O2™ analyzer.

Analyzing Scuba Regulator Samples

Illustrates the direct method for analyzing scuba regulator samples with the Pro O2™ analyzer.

Analyzing Gas with Pro O2 Remote

Details the process for analyzing gas samples using the Pro O2 Remote™ model.

5.0 Maintenance

Analyzer Care and Battery Replacement

Covers general analyzer cleaning, handling, and the steps for battery replacement.

Battery Replacement for Panel Mount Analyzers

Provides specific instructions for replacing the battery in panel mount analyzer units.

Oxygen Sensor Replacement Procedures

Details safety precautions, handling, and step-by-step procedures for replacing oxygen sensors.

6.0 Spares and Accessories

Oxygen Sensors and Flow Accessories

Lists available replacement oxygen sensors, flow restrictors, and regulators with part numbers.

Tee Adapter and Troubleshooting

Describes the Tee Adapter for Pro O2 Remote™ and provides a troubleshooting guide.

Appendix

Calibration Correction for Temperature and Humidity

Explains how to use a chart to determine calibration correction values based on ambient conditions.

NUVAIR Pro O2 and Pro O2 Remote Warranty

Warranty Terms and Return Policy

Outlines the limited warranty coverage, pro-rated terms, and the return service procedure.

Nuvair Pro O2 Specifications

General IconGeneral
Sensor TypeElectrochemical
DisplayLCD
Oxygen Range0-100%
Oxygen Accuracy±1%
Operating Temperature32°F to 122°F (0°C to 50°C)
Power9V battery