6 Steam Sterilisation
Conditioning Instruction BIOM® 5 (G/55185/0818/en) 15 / 26
6 Steam Sterilisation
6.1 Prior to Steam Sterilisation
Attention
Improper cleaning/disinfection is a health risk.
After cleaning/disinfecting the components, check whether they are
macroscopically clean, i.e. free from visible soiling, corrosion or
damage.
If impurities are found: Clean/disinfect the components again.
If corrosion or damage is found: Do not use corroded or damaged
components again for a surgical procedure.
6.1.1 Function Check Prior to Steam Sterilisation
You must check the moving parts of the BIOM® 5 prior to packaging and
steam-sterilising them to ensure their function.
Check all moving parts of the BIOM® 5.