DIGITAL
ULTRASOUND
WATTMETER
UPM-DT-50SP
USER MANUAL

Questions and Answers

  • C
    Chris HammondAug 14, 2025
    Why does the measurement fluctuate on my Ohmic UPM-DT-50SP?
    • S
      samanthadorseyAug 14, 2025
      The meter is very sensitive and affected by minor air drafts or vibrations. To resolve this, place the meter on a solid surface and shield it from any air drafts.
  • D
    Donna DawsonAug 18, 2025
    What to do if my Ohmic Measuring Instruments meter will not zero or tare?
    • T
      Tanner WongAug 18, 2025
      If your meter won't zero, it might be because the degassed water has absorbed too much air. Refill it with fresh degassed water. If the issue persists, contact Ohmic Instruments for Calibration and/or Repair.
  • R
    randallrichardAug 21, 2025
    Why does the measurement drift over time on my Ohmic UPM-DT-50SP?
    • S
      sonya59Aug 21, 2025
      The measurement may drift because the tank water has absorbed too much air. Replace the tank water with fresh, room temperature degassed water. Alternatively, bubbles may have formed on the conical target or transducer head. Wipe any bubbles from the surface of the conical target (including the bottom) and the transducer head.
  • R
    Robert ReynoldsAug 25, 2025
    What to do if my Ohmic UPM-DT-50SP display error Err 8.3?
    • T
      Terry DodsonAug 26, 2025
      Error 8.3 indicates an overload. Remove any excess force on the meter.
  • B
    Brittany LoveAug 28, 2025
    What to do if my Ohmic UPM-DT-50SP Measuring Instruments meter does not respond to the test weight or transducer power?
    • P
      psmithAug 28, 2025
      The Conical Target Assembly and/or the Target Support Tube might be rubbing or binding. Verify that the Conical Target Assembly and its Support Tube are correctly positioned.
  • H
    heather03Aug 31, 2025
    How to fix Err 8.1 on Ohmic Measuring Instruments?
    • L
      limariaAug 31, 2025
      Error 8.1 indicates a load during power on. Remove the conical target assembly and cycle power.
  • K
    Karen JohnsonSep 4, 2025
    What does Err 8.5 mean on Ohmic Measuring Instruments and how to fix it?
    • S
      Scott RitterSep 4, 2025
      Error 8.5 indicates that the tare is out of range. Remove any excess force on the meter.
  • C
    Cathy MarshallSep 7, 2025
    What does error Err 8.2 mean on my Ohmic UPM-DT-50SP Measuring Instruments?
    • D
      Derek PriceSep 7, 2025
      Error 8.2 means the weight is below the power-on zero range. Place the meter on a level surface and cycle power.
  • E
    evancalhounSep 12, 2025
    What to do if my Ohmic Measuring Instruments display error Err 8.4?
    • S
      Stephanie FisherSep 12, 2025
      Error 8.4 indicates an under load condition. Place the meter on a level surface and cycle power.

Summary

Safety and Operational Guidelines

User and Service Cautions

Covers critical warnings and cautions for proper use, modification, service, environment, and inspection.

Handling and Reference Information

Liquid Handling and Cleaning

Important precautions regarding liquid exposure and cleaning procedures for the device.

Symbols and Abbreviations

Explains common symbols and abbreviations used in the manual and on the device.

Disclaimers and Contact Information

Product Use Disclaimers

Outlines limitations of responsibility regarding product use, modifications, and potential inaccuracies.

Company Contact Details

Provides contact information for Ohmic Instruments, including address, phone, and website.

Introduction to the Ultrasound Wattmeter

Product Overview and Principle of Operation

Introduces the UPM-DT-50SP Ultrasound Power Meter and explains its radiant force measurement principle.

Test Setup Procedures

Water as a Measurement Medium

Explains the importance of water as a test medium and details the degasification process for accurate measurements.

Conical Target Installation

Step-by-step guide for correctly installing the conical target assembly into the test tank.

Transducer Placement and Positioning

Instructions on how to properly position and secure the transducer head relative to the conical target.

Meter Layout and Controls

Front Panel Overview

Details the function of each button, the digital display, and the stability indicator on the front panel.

Rear and Bottom Panel Features

Identifies the communications port, power input, and battery compartment located on the rear and bottom of the meter.

Performing Measurements

Measurement Testing Steps

Outlines the step-by-step process for conducting tests and recording measurements with the UPM-DT-50SP.

Data Output and Communication

Print Output and Format

Describes the print port setup, output format, and how unstable readings are indicated.

RS232 Communication Commands

Details the RS232 interface parameters and lists the available communication commands for data retrieval.

Troubleshooting Common Issues

Symptom-Cause-Solution Guide

Provides a guide to identify common symptoms, their possible causes, and recommended solutions for the UPM-DT-50SP.

Interpreting Error Codes

Explains specific error codes (Err 8.1-8.5) and their corresponding possible solutions.

Manual and Warranty Information

Manual Revision History

Lists the revisions made to the user manual, including origination and updates.

Product Limited Warranty Terms

Details the warranty coverage, exclusions, and remedies provided by Ohmic Instruments for the product.

Technical Specifications

Ultrasonic Power Measurement Specs

Lists the performance specifications related to ultrasonic power measurement, including range, resolution, and accuracy.

Electrical and Physical Specifications

Outlines the electrical requirements, display, construction, size, weight, and operating/storage ranges.

Additional Notes

General Notes

Contains general notes or remarks related to the operation or use of the UPM-DT-50SP.

Ohmic UPM-DT-50SP Specifications

General IconGeneral
Accuracy±0.5% of reading ±1 digit
Resolution0.1 μΩ to 0.01 Ω
Operating Temperature0°C to 50°C
Dimensions96 x 48 x 100 mm
WeightApprox. 300 g
Storage Temperature-10°C ~ 60°C