42
Cleaning and disinfection
2. Check that no residual rinse water remains on the HF cable.
8.3 Manual disinfection
Clean the product before disinfection to remove patient
debris. Otherwise, disinfection will be insufcient. Refer to the
section “Manual cleaning” on page 40 for detailed cleaning
instructions.
Manual disinfection is not possible for the monopolar HF cables
A00010A, A00011A, A00012A, A0355, A0391, A0392.
Only use automated cleaning and thermal disinfection for these
products. Refer to the section “Automated cleaning and thermal
disinfection” on page 43.
8.3.1 Validated disinfectant
Olympus has validated manual disinfection of the product with
the following disinfectant:
Cidex Activated Dialdehyde solution manufactured by
Johnson & Johnson
45 minutes immersion time at 25 °C (77 °F)
The instructions in this chapter are based on the validated
disinfectant as specied above.
If other disinfectants are used, select immersion times,
concentrations and temperatures of the solutions as instructed by
the manufacturer of the disinfectant.
8.3.2 Manually disinfecting the product
A complete manual disinfection cycle consists of 3 steps:
Disinfection
Rinsing
Drying