INSTRUCTIONS
This instruction manual is for the Olympus Reflected Fluorescence System. To ensure the safety,
obtain optimum performance and to familiarize yourself fully with the use of this system, we
recommend that you study this manual thoroughly before operating the microscope. Retain this
instruction manual in an easily accessible place near the work desk for future reference.
A X 7 8 8 8
BX3-URA
BX3-RFAS
U-LH100HGAPO
U-LH100HG
U-RFL-T
U-AN-2
BX3-25ND6
BX3-25ND25
REFLECTED
FLUORESCENCE
SYSTEM

Other manuals for Olympus U-RFL-T

Questions and Answers

  • M
    Mary MeyerAug 15, 2025
    Why can't I see anything through my Olympus Microscope, even though the burner is on?
    • E
      erin24Aug 15, 2025
      If the burner is on but the image from your Olympus Microscope is dark, first ensure the shutter is open. Then, check if the ND filter is engaged and remove it if necessary. Verify that the fluorescence mirror unit is correctly engaged and matches the specimen. Also, make sure that the aperture and field iris diaphragms are centered and fully enlarged.
  • L
    Luis ValdezAug 18, 2025
    What to do if the field of view on my Olympus Microscope is obscured?
    • A
      Amber KellyAug 18, 2025
      If your Olympus Microscope's field of view is obscured, ensure the objective and fluorescence mirror unit are correctly engaged in the light path. The revolving nosepiece and ND slider should click into place properly. Also, check that the field iris diaphragm is not set too small; center the aperture iris diaphragm and field iris diaphragm, then fully enlarge the field iris diaphragm. Finally, ensure the mercury burner is properly centered.
  • S
    sadamsAug 22, 2025
    Why is the mercury burner flickering on my Olympus Microscope?
    • V
      vcastroAug 22, 2025
      If the mercury burner of your Olympus Microscope flickers or appears dark, wait at least 10 minutes after ignition for it to stabilize. If the issue persists and the hour counter indicates 300 hours (USH-103OL), replace the mercury burner. Also, ensure the burner is centered on the optical axis.
  • J
    Juan RussoAug 25, 2025
    How to improve the image quality of my Olympus Microscope?
    • J
      Jessica JenkinsAug 26, 2025
      If the image from your Olympus Microscope appears low quality, not sharp, or poor in contrast, start by thoroughly cleaning the objective and filter for any dirt or dust. Also, verify that the fluorescence mirror unit matches the specimen you are viewing.
  • P
    Phillip SantosAug 28, 2025
    What causes dark spots in the field of my Olympus Microscope?
    • C
      ckhanAug 28, 2025
      If your Olympus Microscope field contains dark, spot-like areas, the burner or the burner side of the collector lens may be dirty or dusty. Clean these components to resolve the issue.
  • M
    Mario WilsonAug 31, 2025
    Why won't the mercury burner light on my Olympus Microscope?
    • D
      dfitzpatrickAug 31, 2025
      If the mercury burner on your Olympus Microscope does not light even when the main switch is set to “I” (ON), first, ensure all connectors are firmly connected. Then, check that the mercury burner is properly attached. Verify that the safety device in the lamp housing is correctly set up. If auto ignition is malfunctioning, set the main switch of the power supply unit to “ ” (OFF) then “I” (ON) again, repeating as needed.
  • C
    Carolyn GilmoreSep 3, 2025
    What to do if the Olympus Power Supply main switch is on but the mercury burner doesn't ignite?
    • K
      kennethrandolphSep 3, 2025
      If the main switch indicator lights up but the mercury burner on your Olympus Power Supply fails to ignite, there could be several reasons: * The connectors might be improperly connected. Ensure they are correctly connected. * The burner might not be installed. Make sure the burner is properly installed. * The lamp housing interlock mechanism might be activated. Tighten the burner socket clamping screw. * The auto ignition could be malfunctioning. Try turning the main switch on the power supply to “O” (OFF) and then back to “I” (ON). Repeat this process if necessary.
  • W
    websterkathrynSep 7, 2025
    Why is the Olympus Power Supply burner flickering or dark?
    • R
      Richard LeonSep 8, 2025
      If the burner on your Olympus Power Supply flickers or appears dark, it could be due to the following reasons: * Insufficient warm-up time: Allow the burner to warm up for at least 10 minutes after turning it on. * Expired burner life: If the hour counter reading exceeds 200 hours for USH102D or 300 hours for HBO103W/2, replace the mercury burner.
  • J
    Jacob BarnesSep 10, 2025
    How to fix an Olympus Power Supply when the main switch LED doesn't light up?
    • T
      Tracy JacksonSep 10, 2025
      If the main switch LED on your Olympus Power Supply does not light up, the power cord may not be connected properly. Ensure that the power cord is correctly connected.

Summary

SAFETY PRECAUTIONS

NOMENCLATURE

REFLECTED FLUORESCENCE OBSERVATION PROCEDURE

USING THE CONTROLS

General Precautions for Observation

General precautions for safe and optimal operation of the reflected fluorescence system.

Turning the Power Supply Unit On

Step-by-step instructions for powering on the lamp housing and ensuring stable illumination.

Centering the Mercury Burner

Procedure to align the mercury burner poles for proper light incidence and optimal image quality.

SIMULTANEOUS FLUORESCENCE OBSERVATIONS

TROUBLESHOOTING GUIDE

SPECIFICATIONS

ASSEMBLY

PROPER SELECTION OF THE POWER SUPPLY CORD

Olympus U-RFL-T Specifications

General IconGeneral
BrandOlympus
ModelU-RFL-T
CategoryPower Supply
LanguageEnglish