Operation and Care
The device should be cleaned with a soft and dry cloth, or a soft and moistened cloth
and neutral soap.
(Cleaning and disinfecting)
Observe the rules below when cleaning or disinfecting parts. Failure to observe these
rules may result in damage, inefficient nebulization or infection. For the instructions,
refer to “Cleaning and disinfecting” section.
• Clean and disinfect the nebulizer kit, mask, mouthpiece or nosepiece before using
them when as follows:
- the first time after purchase
- if the device has not been used for a long period of time
- if more than one person uses the same device
• Be sure to wash or wipe the parts after use, and ensure that they are thoroughly
disinfected and dried, and stored in a clean location.
• Do not leave the cleaning solution in the parts.
Error Messages
Problem
Cause
Remedy
Nothing happens when the
power switch is turned.
Is the power plug plugged into a
power outlet?
Check that the plug is inserted
in a power outlet. Unplug then
reinsert the plug if necessary.
No nebulization or low
nebulization rate, when the
power is on.
Is there medication in the
medication tank?
Add the correct amount of
medication to the medication
tank.
Is there too much/little
medication in the medication
tank?
Is the vaporizer head missing or
not assembled correctly?
Attach the vaporiser head
correctly.
Is the nebulizer kit assembled
correctly?
Assemble the nebulizer kit
correctly.
Is the nozzle blocked?
Make sure that the nozzle is
free of blockages.
Is the nebulizer kit tilted at a
sharp angle?
Make sure that the nebulizer kit
is not tiled at an angle of more
than 45 degrees.
Is the air tube connected
correctly?
Make sure that the air tube is
correctly connected to the
compressor and nebulizer kit.
Is the air tube folded or
damaged?
Make sure that the air tube does
not contain kinks.
Is the air tube blocked?
Make sure that the air tube is
free of blockages.
Is the air filter dirty? Replace the air filter with a new
one.
The compressor is abnormally
loud.
Is the air filter cover attached
correctly?
Attach the air filter cover
correctly.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron A3 Complete and is the answer not in the manual?

Omron A3 Complete Specifications

General IconGeneral
Noise level65 dB
Product colorBlue, White
Product designTabletop
Built-in displayNo
Maximum fill volume12 ml
Nebulizer technology-
Designed for childrenYes
Power source typeAC