What to do if my Omron NE-C801 Respiratory Product is not nebulizing well?
- GGregg PorterJul 30, 2025
If your Omron Respiratory Product has a low nebulization rate, several factors could be responsible:* Ensure the medication tank has the correct amount of medication.* Verify the vaporizer head and nebulizer kit are correctly assembled.* Check for any blockages in the nozzle or air tube.* Make sure the nebulizer kit isn't tilted more than 45 degrees.* Confirm the air tube is properly connected to both the compressor and nebulizer kit and is free of kinks.* Replace the air filter if it's dirty.