9.Technical data
Technical data for the OMRON NE-C801KD Compressor with the OMRON V.V.T.
(Virtual Valve Technology) Nebulizer kit:
Result of cascade impactor measurements for particle size**
* Measured by OMRON HEALTHCARE Co., Ltd.
**Independently measured by Prof. Dr. Hiroshi Takano, Department of Chemical
Engineering and Materials Science, Faculty of Science and Engineering, Doshisha
University, Kyoto, Japan, according to EN 13544-1:2007
Notes:
Performance may vary with medication such as suspensions or high viscosity.
See medication supplier’s data sheet for further details.
See web site of OMRON HEALTHCARE EUROPE to update technical information.
URL: www.omron-healthcare.com
Particle Size:
**MMAD Approx. 3μm
MMAD = Mass Median Aerodynamic Diameter
Medication Tank Capacity:
7ml maximum
Appropriate Medication
Quantities:
2ml minimum - 7ml maximum
Sound:
*Noise level (at 1 m distance) 46 dB
Nebulization Rate:
*Approx. 0.3ml/min (by weight loss)
Aerosol Output:
**0.47ml (2ml, 1%NaF)
Aerosol Output Rate:
**0.06ml/min (2ml, 1%NaF)
0
10
20
30
40
50
60
70
80
90
100
1000110.1
Cumulative % particle mass of sodium fluoride under size
Particle Size Dp (μm)
Individual
Mean
MMAD 3.0 μm
Cumulative Undersize %