Cause 1
The probe was inappropriately placed as indicated
by measurement method, example: oral,
axillary, and rectally.
Cause 2
Temperature did not register because of poor
contact or placement.
Instructions
1. Check correlation between selected mode icon
and actual measurement location used.
2. Turn the unit OFF, wait one minute, Turn the unit
on and retake the temperature using the proper
technique. (see page 9).
Cause
Unit malfunction.
Instructions
Stop using the thermometer – contact Omron
Customer Service 1-800-634-4350
Cause
Temperature was not taken within 1 minute
after the thermometer reached the ready state.
Instructions
Turn the unit OFF. Turn it ON again after one minute
to retake the temperature (as described on page 9).
Cause
Low battery.
Instructions
The battery has enough remaining life
for approximately 10 more measurements.
Cause
The battery is very low
Instructions
No measurement can be taken.
Replace the batter with a new one.
12
DISPLAY MESSAGES
MC-600 English 5/20/02 2:02 PM Page 13