Description of operating principle
How the nebulizer kit works:
The medication that is pumped up
through the medication channel is
mixed with compressed air which is
generated by a compressor pump.
The compressed air mixed with
medication is turned into fine
particles and sprayed when in
contact with the baffle.
Nebulizer Kit with Valve Technology
The X105 Advanced nebulizer is equipped with a breath-enhanced valve
system
which allows the maximum amount of medication to be inhaled by the
patient. It optimizes the flow of medication during inhalation and reduces
medication loss during exhalation. By synchronizing the flow of aerosol
with the breathing, the therapeutic effect of the inhaled medication is
optimized.
Note: Only using the mouthpiece enables the synchronized functioning
of the nebulizer!
Inhalation phase:
During inhalation, the inhalation
valve (A) opens and air enters the
nebulizer. The air is mixed with the
generated aerosol (B) and inhaled
through the mouthpiece by the
patient (C). In this phase the
exhalation valve
(D) remains
closed preventing aerosol
from being wasted.