27
TROUBLESHOOTING AND DETAILED INFORMATION ABOUT YOUR SYSTEM
Electrical and safety standards
This meter complies with CISPR 11: 2003, Class B (Radiated Only). Emissions of the energy used are low and not likely to cause interference
in nearby electronic equipment. The meter has been tested for immunity to Level 3 electrostatic discharge as specied in IEC 61000-4-2.
This meter has been tested for immunity to radio frequency interference over the frequency range 80 MHz to 2.5 GHz at 3V/m as specied in
IEC 61000-4-3.
Degree of protection rating: IP32
CAN/CSA C22.2 61010-1:04, UL 61010-1:04, IEC 61010-1 and IEC 61010-2-101.
Guarantee
LifeScan guarantees that the OneTouch® Ultra® 2 Meter will be free of defects in material and workmanship for three years, valid from the
date of purchase. The guarantee extends only to the original purchaser and is not transferable.
Technical specications
Reported result range 1.1–33.3 mmol/L
Calibration Plasma-equivalent
Sample Fresh capillary whole blood
Test time 5 seconds
Assay method Glucose oxidase biosensor
Meter power source One replaceable 3.0 Volt CR 2032
lithium battery (or equivalent)
Backlight power source One replaceable 3.0 Volt CR 2032
lithium battery (or equivalent)
Unit of measure mmol/L
Memory 500 blood glucose or control solution
test results
Automatic shuto 2 minutes after last action
Size 7.92 x 5.72 x 2.29 cm
Weight Approximately 42.5 grams,
with batteries
Operating ranges Temperature: 6–44°C
Relative humidity: 10–90%
Altitude: up to 3048 meters
Hematocrit: 30–55%
Battery ratings 2 x 3.0 V d.c., 60 mA
(2 x CR 2032 batteries)
direct current
C US