IL Pass Threshold is a wavelength-specic setting which controls
the pass/fail criteria for insertion loss measurements through the
front panel display.
IL Precision is a global setting which controls the displayed
precision of insertion loss measurements.
IL Dwell Time is a global setting which controls how long the unit
waits for insertion loss measurements to settle after switching
from a different wavelength, channel, or different kinds of
measurements.
RL Pass Threshold is a wavelength-specic setting which
controls the pass/fail criteria for return loss measurements
through the front panel display.
RL Dwell Time is a global setting which controls how long the
unit waits to perform Return Loss measurements to settle after
switching from a different wavelength, channel, or different kinds
of measurements.