What to do if I get error code xx on my Orion Diagnostica QuickRead go CRP Laboratory Equipment?
- KKelly MeltonAug 19, 2025
Contact your local distributor or call customer service and give the error code.
What to do if I get error code xx on my Orion Diagnostica QuickRead go CRP Laboratory Equipment?
Contact your local distributor or call customer service and give the error code.
Why am I getting an unexpected high result with my Orion Diagnostica Laboratory Equipment?
An unexpected high result could be due to several reasons: * Too large sample volume: Perform a new test, making sure the capillary is properly filled and wiping away any excess sample. * Too low buffer volume: Perform a new test, ensuring the liquid surface is between the two lines marked on the cuvette. * The cuvette is dirty: Perform a new test, avoiding touching the clear flat surfaces in the lower part of the cuvette.
Why am I getting an unexpected low result with my Orion Diagnostica QuickRead go CRP?
If you're seeing an unexpectedly low result, it might be due to a too small sample volume. To fix this, perform a new test, ensuring the capillary is completely filled and avoiding air bubbles.
What to do if my Orion Diagnostica QuickRead go CRP test is cancelled due to instrument failure?
Perform a new test. If this message appears often, contact your local distributor or call customer service.
Why am I getting unexpected low or high results with my Orion Diagnostica Laboratory Equipment?
Unexpected low or high results may occur due to: * Using components from different kit lots. Perform a new test using reagents from the same kit lot. * Incorrect reagent storage. Perform a new test, ensuring reagents are stored according to instructions. * Incorrect sample volume: plasma/serum/control sample. Check that the used sample volume (20 or 12 µl) matches the set plasma/serum sample volume on the instrument.
What to do if my Orion Diagnostica Laboratory Equipment test is cancelled because the blank is too high?
Test the same cuvette again. The blanking process may not have been completed, or the sample may contain interfering substances. In the latter case, the test cannot be completed.
What to do if my Orion Diagnostica QuickRead go CRP test is cancelled due to an error in reagent addition?
Perform a new test, ensuring that the cap is properly closed, as there may have been a problem during reagent addition.
What to do if my Orion Diagnostica Laboratory Equipment says measurement prohibited due to kit lot expired?
Discard the expired kit lot and use a new one.
What to do if my Orion Diagnostica QuickRead go CRP Laboratory Equipment shows no result and is unable to do hematocrit correction?
Test the same cuvette again. If the error message appears again, perform a new test.
What to do if my Orion Diagnostica Laboratory Equipment says measurement prohibited because the cuvette temperature is too high?
Let the cuvette cool down to room temperature (18...25ºC). Test the same cuvette again.
General safety guidelines for handling reagents and samples.
Step-by-step instructions for performing the CRP assay.
Instructions for inserting the cuvette and obtaining results from the instrument.
Steps for analyzing control samples using the instrument.
Lists instrument error codes and their corrective actions.
| Measurement Range | 0.5 - 200 mg/L |
|---|---|
| Method | Immunoturbidimetric assay |
| Sample Volume | 10 μL |
| Battery Life | up to 8 hours |
| Type | Point-of-care testing (POCT) device |
| Sample Type | Whole blood, serum, plasma |
| Test Time | 3 minutes |
| Calibration | Automatic |
| Display | Color LCD |
| Data Output | USB, Bluetooth |
| Connectivity | Bluetooth |