QuikRead go
®
CRP
Assay Procedure • Testverfahren
1 2
3 4
5 6
7 8
Measure
RESULT
Patient ID:
Test:
XXXXXXXX
Measurement time:
CRP:
CRP
2013-06-11 12:19
mg/l
20
Exit Print New measurement
Result info
Choose Result info to view result information.
Remove cuvette to perform a new measurement.
i

Questions and Answers

  • B
    bsilvaAug 19, 2025
    What to do if I get error code xx on my Orion Diagnostica QuickRead go CRP Laboratory Equipment?
    • K
      Kelly MeltonAug 19, 2025
      Contact your local distributor or call customer service and give the error code.
  • T
    Theresa WhiteAug 21, 2025
    Why am I getting an unexpected high result with my Orion Diagnostica Laboratory Equipment?
    • L
      Lacey JacksonAug 21, 2025
      An unexpected high result could be due to several reasons: * Too large sample volume: Perform a new test, making sure the capillary is properly filled and wiping away any excess sample. * Too low buffer volume: Perform a new test, ensuring the liquid surface is between the two lines marked on the cuvette. * The cuvette is dirty: Perform a new test, avoiding touching the clear flat surfaces in the lower part of the cuvette.
  • C
    Chad RojasAug 23, 2025
    Why am I getting an unexpected low result with my Orion Diagnostica QuickRead go CRP?
    • T
      Taylor ThompsonAug 23, 2025
      If you're seeing an unexpectedly low result, it might be due to a too small sample volume. To fix this, perform a new test, ensuring the capillary is completely filled and avoiding air bubbles.
  • A
    Anne HaleAug 26, 2025
    What to do if my Orion Diagnostica QuickRead go CRP test is cancelled due to instrument failure?
    • L
      Lori SuarezAug 26, 2025
      Perform a new test. If this message appears often, contact your local distributor or call customer service.
  • S
    Susan StewartAug 28, 2025
    Why am I getting unexpected low or high results with my Orion Diagnostica Laboratory Equipment?
    • D
      Donald ShermanAug 28, 2025
      Unexpected low or high results may occur due to: * Using components from different kit lots. Perform a new test using reagents from the same kit lot. * Incorrect reagent storage. Perform a new test, ensuring reagents are stored according to instructions. * Incorrect sample volume: plasma/serum/control sample. Check that the used sample volume (20 or 12 µl) matches the set plasma/serum sample volume on the instrument.
  • V
    Vanessa SullivanAug 30, 2025
    What to do if my Orion Diagnostica Laboratory Equipment test is cancelled because the blank is too high?
    • T
      Timothy VincentAug 30, 2025
      Test the same cuvette again. The blanking process may not have been completed, or the sample may contain interfering substances. In the latter case, the test cannot be completed.
  • A
    Amanda PetersonSep 2, 2025
    What to do if my Orion Diagnostica QuickRead go CRP test is cancelled due to an error in reagent addition?
    • M
      Matthew RichardsonSep 2, 2025
      Perform a new test, ensuring that the cap is properly closed, as there may have been a problem during reagent addition.
  • M
    Michael HopkinsSep 4, 2025
    What to do if my Orion Diagnostica Laboratory Equipment says measurement prohibited due to kit lot expired?
    • S
      Stephanie DennisSep 4, 2025
      Discard the expired kit lot and use a new one.
  • G
    Gina DawsonSep 6, 2025
    What to do if my Orion Diagnostica QuickRead go CRP Laboratory Equipment shows no result and is unable to do hematocrit correction?
    • D
      Destiny GonzalezSep 6, 2025
      Test the same cuvette again. If the error message appears again, perform a new test.
  • J
    Jennifer CarterSep 8, 2025
    What to do if my Orion Diagnostica Laboratory Equipment says measurement prohibited because the cuvette temperature is too high?
    • E
      Eric SmithSep 8, 2025
      Let the cuvette cool down to room temperature (18...25ºC). Test the same cuvette again.

Summary

Principles of the Procedure

Reagents

Warnings and Precautions

Health and Safety Information

General safety guidelines for handling reagents and samples.

Sample Collection and Preparation

Procedure

Assay Procedure

Step-by-step instructions for performing the CRP assay.

Analysing the Sample (Fig. 6-8)

Instructions for inserting the cuvette and obtaining results from the instrument.

Analysing the Control (Fig. 7-8)

Steps for analyzing control samples using the instrument.

Results Interpretation

Limitations of the Procedure

Performance Characteristics

Disposal

Troubleshooting

Error Messages

Lists instrument error codes and their corrective actions.

Orion Diagnostica QuickRead go CRP Specifications

General IconGeneral
Measurement Range0.5 - 200 mg/L
MethodImmunoturbidimetric assay
Sample Volume10 μL
Battery Lifeup to 8 hours
TypePoint-of-care testing (POCT) device
Sample TypeWhole blood, serum, plasma
Test Time3 minutes
CalibrationAutomatic
DisplayColor LCD
Data OutputUSB, Bluetooth
ConnectivityBluetooth