– 47 –
d) When the measurement has been concluded, the measured sys-
tolic and diastolic blood pressure values, as well as the pulse will
be displayed.
e) The appearance of this symbol
signifies that an irregular heart-
beat was detected. This indicator is only a caution. It is important
that you be relaxed, remain still and do not talk during measure-
ments.
We recommend contacting your physician if you see this in-
dicator frequently.
f) The measurement results are displayed until you switch the devi-
ce off. If no button is pressed for 30 seconds, the device switches
off automatically.
There is a 15 second resting time in-between each measurement.
If one of the measurements causes an error message, it will be re-
peated one more time. If any additional error occurs, the measu-
rement will be discontinued and an error code will be displayed.
h) Movement error symbol (
)
The Movement Error Symbol ( ) is displayed if you move your
body during the measurement. Please remove the cuff, and wait
2-3 minutes. Reapply the cuff and take another measurement.
Patient Position:
1) Comfortably seated;
2) Legs uncrossed;
3) Feet flat on the floor;
4) Back and arm supported;
5) Middle of the CUFF at the level of the right atrium of the heart.
1. Recommendation that the PATIENT relax as much as possible and
not talk during the measurement PROCEDURE.
2. Recommendation that 5 min should elapse before the first re-
ading is taken.