Guidance and manufacturer’s declaration – electromagnetic immunity
The “Digital Blood Pressure Monitor ORO-N6 BASIC” is intended for use in the
electromagnetic environment specified below. The customer or the user of the “Digital
Blood Pressure Monitor ORO-N6 BASIC” should ensure that it is used in such an
environment.
Immunity test
IEC 60601 test
level
Compliance
level
Electromagnetic
environment –
guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz
to 80 MHz
3V/m
80 MHz to
2,5 GHz
• 3V
• 3V/m
Portable and mobi-
le RF communica-
tions equipment
should be used no
closer to any part
of the
“Digital Blood
Pressure Monitor
ORO-N6 BASIC”,
including cables,
than the
recommended se-
paration distance
calculated from
the equation
applicable to the
frequency of the
transmitter.
Recommended se-
paration distance
• d=1.2√P
d=1.2√P
80MHz to
800MHz
d=2.3 √P
800MHz
to 2.5 GHz