1 An additional 5-second delay may occur if signal quality drops and then recovers as the Sensor reinitializes the oxygen readings
2 These two alarms are mutually exclusive. The low priority alarm sounds first aer 30 seconds and will escalate to a medium priority alarm if the issue is not
resolved. These alarms can only be cleared with valid data and no movement.
16.4 Alarm Signal Generation Delay (Time from alarm condition to alarm signal)
The Base Station controls the alarms and there is no measurable delay in signal generation within the Base Station or in sending the data to the mobile device.
There is less than a 3-second delay in signal generation by the App when connected to the Base Station.
16.5 Alarm Thresholds and Volume
Alarm Type
Factory Default Value
Adjustment Range
Step Value
Low Oxygen Saturation Alarm Limit
High Oxygen Saturation Alarm Limit
Low Pulse Rate Alarm Limit
High Pulse Rate Alarm Limit
Alarm Sound Pressure Level
85% SpO2
OFF (100% when enabled)
75 BPM
200 BPM
L
ow priority
:
55-
6
0 dBA
H
igh
/
Medium Priority
:
6
0-
6
5 dBA
*
(
*
measured at 1 meter)
50-
9
5%
85-100%
,
OFF
30-110 BPM
75-275 BPM
N
ot ad
j
ustable
1% SpO2
1 BPM
N
ot applicable
Alarm thresholds should only be altered by a healthcare professional. These values can be modified through the Modify Alarm Settings menu in the Owlet
C
are
+
App (see section 8.5). Alarm volumes are fixed and cannot be ad
j
usted.
! N
OT
E
.
The alarm limit settings are retained in the event of a
power interruption.
! N
OT
E
.
W
ithout the App
,
the alarm thresholds cannot be
modified.