-86°C Ultra-Low Freezers
Description
Model Number MDF-U33V-PA MDF-U53VA-PA MDF-U56VC-PA MDF-U76VC-PA MDF-U76VA-PA
Interior Volume
11.8 cu.ft.
334 L
18.3 cu.ft.
519 L
18.6 cu.ft.
526 L
25.7 cu.ft.
728 L
25.7 cu.ft.
728 L
Area Footprint, Nominal
6.25 sq.ft.
0.58 m
2
7.24 sq.ft.
0.68 m
2
7.24 sq.ft.
0.68 m
2
9.51 sq.ft.
0.88 m
2
9.51 sq.ft.
0.88 m
2
Electrical, 60 Hz 115V, AC, 15 amp 115V, AC, 20 amp 208/230V, AC, 15 amp 208/230V, AC, 15 amp 115V, AC, 20 amp
Patented VIP
®
Construction
Insulated Inner Door Steel-framed, high impact plastic with foam-in place insulation
Inner Door Conguration 2 2 (4 optional) 2 (4 optional) 2 (4 optional) 2 (4 optional)
Door Gaskets Multi-point compression gaskets
Condenser Filter, Easy Cleaning Electrostatic lter standard, front accessible for easy access, no tools required
Rounded Interior Corners Simplify Cleaning
Shelf Brackets Incrementally adjustable
Casters High-impact 4 with levelling feet
Seismic Restraints Standard; hard connection to facility wall
Access Ports Multiple ports allow use of injection tubes, probes, leads, instrumentation
Dimensions & Weight
Interior (w x f-b x h)
19.3” x 23.6” x 44.9”
490 x 600 x 1140 mm
24.8” x 23.6” x 54.3”
630 x 600 x 1380 mm
24.8” x 23.6” x 55.1
630 x 600 x 1400 mm
34.2” x 23.6” x 55.1
870 x 600 x 1400 mm
34.2” x 23.6” x 55.1
870 x 600 x 1400 mm
Exterior (w x f-b x h)
26.4” x 34.1” x 73.2”
670 x 867 x 1860 mm
30.3” x 34.4” x 78.3”
770 x 870 x 1990 mm
30.3” x 34.4” x 78.3”
770 x 870 x 1990 mm
39.8” x 34.4” x 78.3
1010 x 870 x 1990 mm
39.8” x 34.4” x 78.3
1010 x 870 x 1990 mm
Net Weight, Empty
562 lbs
255 kg
660 lbs
299 kg
672 lbs
305 kg
805 lbs
365 kg
820 lbs
373 kg
Refrigeration System, HFC Refrigerants, Panasonic Cool Safe High Performance Compressors
High Stage 1 HP ½ HP 1 HP 1 HP 1 HP
Low Stage ¾ HP 1 HP 1½ HP 1½ HP 1½ HP
Voltage Booster, Built-In Not Available Not Available Standard Standard Not Applicable
Sound Abatement Standard Standard Standard Standard Standard
Refrigeration System, HFC Refrigerants, Panasonic Cool Safe High Performance Compressors
Power, AC, 1 Phase 115V 115V 208-230V 208-230V 115V
Recommended Breaker, Dedicated 15 amp 20 amp 15 amp 15 amp 20 amp
NEMA Plug / Receptacle 5-15P/5-15R
5-20P/5-20R 6 -15P/6-15R 6-15P/6-15R 5-20P/5-20R
VIP
®
Series Specications
Product conforms to RoHS
(European Restriction of Hazardous
Substance directives)
Panasonic Healthcare Company of North America
1300 Michael Drive, Suite A, Wood Dale, IL 60191
Toll Free USA (800) 858-8442, Fax (630) 238-0074
www.panasonic.com/biomedical
© Panasonic Printed in USA 2012.02 JB
JB201205V2