Appendix
© Copyright 2008 Panoramic Corporation
PC-1000/Laser 1000 Space Requirements
F
Cephalometric
Exposure Time Chart
90 kVP 10 mA
Rare Earth Screens And Film
Seconds
Patient Size:
Small Child
Large Child
Average Adult
Large Adult
Lateral
AP-PA
.5-1.0
.6-1.5
.8-2.0
1.0-2.5
.8-1.5
1.0-2.5
1.5-3.5
2.5-4.0
LASER 1000
Panoramic Corporation
®
PC-1000
PC-1000
The PC-1000/Laser 1000 weighs approximately
515 pounds and is freestanding, requiring no
extra support in the wall or oor.
The factory con guration is shipped with the
control panel mounted on the patient's left
side and the cephalometric arm mounted on
the patient's right side, unless speci ed by the
customer prior to shipping. The control panel
can be easily relocated to the right side at the
time of installation.
Note: The FDA requires that the technique
factors (kVp meter) be viewable during the
exposure.
Physical Dimensions
72" W x 44" D x 91" H
Minimum Working Space
84" W x 48" D x 91" H
72"
35"
23"
44"
33"
91" Maximum