Table of Contents

Questions and Answers

  • S
    Shelby HernandezAug 1, 2025
    What to do if nothing happens when pressing the foot pedal on PAXXO AB Pactosafe Medical Equipment?
    • M
      michael12Aug 1, 2025
      If nothing happens when you press the foot pedal, there could be several reasons: * The Pactosafe might not be connected to a mains socket. Ensure it is properly connected. * The switch might not be turned on. Turn on the switch. * There might be no signal from the foot pedal. Check the connection and for any damage; shorten or replace if necessary. * The Pactosafe might be starting up. The system requires 5 seconds to become operational. * The battery might be discharged. Charge the battery.
  • K
    Kyle YatesAug 6, 2025
    What to do if waste is stuck between the welding bars in PAXXO AB Medical Equipment?
    • J
      John WilliamsAug 6, 2025
      If waste gets stuck between the welding bars, switch the unit OFF and then ON again. The Pactosafe will then return to its start mode.
  • T
    Thomas PerezAug 9, 2025
    Why is the feeding of the foil working but not the welding in PAXXO AB Medical Equipment?
    • E
      Eddie WilsonAug 9, 2025
      The feeding of the foil might work, but the welding doesn't due to: * Something jamming the cog wheels and blocking the welding bars. Remove the item causing the blockage. * Damaged welding bars. Contact the distributor/manufacturer. * Something being wrong inside the unit. Contact the distributor/ manufacturer.
  • C
    Christopher RuizAug 13, 2025
    What to do if the foil is adhering to the welding bars in PAXXO AB Medical Equipment?
    • B
      bstewartAug 14, 2025
      If the foil is adhering to the welding bars, it could be due to damaged welding bars. In this case, contact the distributor/ manufacturer.
  • R
    Ryan MorganAug 18, 2025
    What causes incomplete welding of the foil in PAXXO AB Pactosafe?
    • B
      billallenAug 18, 2025
      Incomplete welding of the foil may be caused by: * Damaged welding bars. Contact the distributor/ manufacturer. * Something wrong inside the unit. Contact the distributor/ manufacturer.
  • M
    melissa60Aug 23, 2025
    Why is the welding of the foil too powerful in PAXXO AB Medical Equipment?
    • B
      Brian WebbAug 23, 2025
      If the welding of the foil is too powerful, it indicates something is wrong inside the unit. Contact the distributor/ manufacturer.
  • D
    David GallagherAug 27, 2025
    What to do if feeding of the foil works but not the welding in PAXXO AB Medical Equipment?
    • T
      Theodore ThomasAug 27, 2025
      If the foil feeds but the welding doesn't work on your PAXXO AB Medical Equipment, first check if something is jamming the cog wheels and blocking the welding bars. If there is a blockage, remove the item causing it. If the welding bars are damaged or something is wrong inside the unit, contact the distributor/manufacturer.

Summary

DECLARATION OF CONFORMITY

1. Introduction

2. Safety instructions

Statement of signs

Explains various warning symbols and their meanings for safe operation.

3. Functional description

4. Technical specification

5. Unpacking; Putting into operation

6. Replacement of foil

7. Disposal of used foil

8. Adjustment of bag length

9. Battery operation (optional)

10. Checking the welding result

11. Checking the safety device

12. Troubleshooting

13. Cleaning and decontamination

14. Replacing the plug-in unit

Removing the plug-in unit

Step-by-step guide for detaching the plug-in unit from the machine.

Fitting a new plug-in unit

Instructions for installing a replacement plug-in unit and connecting it.

15. Disposal

PAXXO AB Pactosafe Specifications

General IconGeneral
BrandPAXXO AB
ModelPactosafe
CategoryMedical Equipment
LanguageEnglish