PTB 311E & PTB 511E Manual 3-in-1 Tablet Testing Instrument
PTAG_29-02860_PTB311E_v2.2E.docx
Technical Data
Parameter
Specification
Display
LED Display showing number of tests and results
Data Entry
Numerical and Functional keys
Standard Force Range
5.0 to approximately 300N
Accuracy
< 1N
Resolution
0.074N (300N model) - 0.1482N (500N model)
Force Settings
Linear speed or linear force increase
Selectable Range
5.0 - 200 N/Second or Millimeter/Minute
Accuracy
< 2% force or < 0.1% speed
Maximum Sample Size
45 mm
Diameter Range
2.0 - 45.0mm
Height/Thickness Range
2.0 - 45.0mm (requires different thickness jaws)
Accuracy
< 0.05 mm (typically 0.02mm)
Resolution
0.0015 mm
Printer
Interface to connect Dot-Matrix or PCL 5 Printer
Interface
RS232 COM port
Calibration Guidance
Built-in calibration procedures the digital load cell
Adjustments
Two point adjustment - zero and 10kg
Calibrations
Multiple point for load cell precision using certified weights
(CAL15/30/50)
Force Detection
Reproducibility
PT-MT Magnetic Tablet
Instrument Housing
Stainless Steel to meet GLP requirements
Bench Space Requirement
L 48cm x W 24cm x H 26cm (without external printer )
Certification
All components certified to USP / EP requirements
CE / EMC Certification
All CE / EMC Certification provided
Validation
All IQ & OQ documents included