KVM100A/200A
5-2
OPERATING INSTRUCTIONS
2. STAND ALONE DIVIDER OPERATION:
1. Make sure connections have been made as described in “Set-Up” section, except that metering
module will not be used or connected.
2. Connect a voltmeter to the BNC output on base of divider with 25’ coaxial cable.
NOTE: Divider has a 10,000 to 1 ratio. At 100kV input, the output voltage will be 10V. The meter
impedance needs to be 10M DC, and 1 M, ≤ 200pF AC for the highest possible accuracy.
Readings will correspond to the type of meter used. AVG will show AVG, RMS will show RMS
etc. Use DC range for DC signals (10 meg input meter).
3. Select desired mode for voltmeter.
4. Activate high voltage source.
5. Measure voltage as required.
6. De-activate high voltage source and ensure, as appropriate, high voltage source is de-energized and
discharged.
7. After ensuring that high voltage source and divider are de-energized and discharged, all leads, and
connections may be disconnected and unit packaged.
3. STAND ALONE METER MODULE OPERATION:
It is not recommended to operate the metering module without the divider. The metering module is not a
stand-alone voltmeter.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Phenix Technologies KVM 100A and is the answer not in the manual?

Phenix Technologies KVM 100A Specifications

General IconGeneral
BrandPhenix Technologies
ModelKVM 100A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals