2-1
SECTION 2: TECHNICAL SPECIFICATIONS
AC/DC KILOVOLTMETER
KVM100 / KVM 200
KVM100 DIVIDER
KVM 200 DIVIDER
HIGH VOLTAGE AC INPUT:
100kVAC maximum
200 KVAC maximum
HIGH VOLTAGE DC INPUT:
100kVDC maximum
200 KVDC maximum
HIGH VOLTAGE PK INPUT:
142kV PEAK (AC)
200 KV peak max.
DIVIDER CAPACITANCE:
200 pF
100 pf
DIVIDER RESISTANCE:
380M
760 m
DIVIDER RATIO:
10,000:1
10,000:1
INSTRUMENTATION: AC / BATTERY POWERED
POWER ADAPTOR:
INPUT: 100-240 Volts AC, 0.4 AMPS, 47-63 HZ
OUTPUT: +15 VDC, 1 AMP
INPUT JACK:
+15 VDC, 1 AMP
BATTERY PACK:
VOLTAGE MEASUREMENT:
Voltage is measured by means of Peak Responding, RMS or AVG
Circuitry. Display is 4 ½ Digit LED with adjustable backlighting
ACCURACY: 1% of Reading from 10% -100% of Range
FUNCTION SELECTIONS:
1. Peak / 2 (Peak responding calibrated to RMS Value)
2. Peak (Peak responding calibrated to Peak Value)
3. RMS (True RMS)
4. AVG (True Average Value, not RMS equivalent)
5. Peak DC (Peak responding including ripple peak
6. Ripple Voltage (RMS value of a DC Ripple voltage)
FREQUENCY RESPONSE:
(SINUSOIDAL WAVEFORM)
Peak / 2, Peak, --Response: DC, 5-1000Hz
RMS, Ripple V --Response: 20 - 1000Hz
AVG --Response: DC, 20-1000Hz
SETTLING TIME:
Peak / 2, Peak --Up to 30 seconds
RMS, Ripple V --Up to 15 seconds
AVG --Up to 5 seconds
*Note: The lower the frequency the longer the settling time.
ENVIRONMENTAL CONDITIONS
10-40°C, Indoor/Outdoor in fair weather
Humidity <95%, non-condensing
Altitude <3000 ft (1000 meters)