KVM300
5-2
OPERATING INSTRUCTIONS
2) STAND ALONE DIVIDER OPERATION:
Make sure connections have been made as described in “Set-Up” section, except that the
meter module will not be used or connected.
Connect a Voltmeter to the BNC output on the base of the divider with the 25’ coaxial cable.
NOTE: The divider has approximately 10,000 to 1 ratio. At 300kV input the output voltage will be
30V. Readings will correspond to the type of meter used. AVG will show AVG, RMS will show
RMS etc. The meter impedance needs to be 10M DC, and 1 M, ≤ 200pF AC for the highest
possible accuracy. Readings will correspond to the type of meter used. AVG will show AVG, RMS
will show RMS etc. Use DC range for DC signals (10MΩ input meter).
Select desired mode for voltmeter.
Activate high voltage source.
Measure voltage as required.
De-activate high voltage source and assure, as appropriate, that high voltage source is de-
energized and discharged.
After assuring that high voltage source and unit are de-energized and discharged, all leads
and connections may be disconnected and unit packaged for transport or storage.
3) STAND ALONE METER MODULE OPERATION:
It is not recommended to operate the meter head without the divider. The meter head is not a
stand alone voltmeter.