EasyManua.ls Logo

PHI VersaProbe - Page 10

Default Icon
11 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
!
f. If used neutralizer during AutoZ, go to “SXI/AutoZ” tab and uncheck
“Neutralizer” and “Ion Gun Neut.”
g. Watcher -> Press Pump Intro so V4 closes and turbo again pumps
the intro chamber.
VIII. Sample Extraction (See Figures 10-11).
1. Summitt-> Image window-> Press Extract. Sample stage will lower, return
to original X, Y and R settings, and then rise to a high Z position.
2. Watcher -> Push Transfer Sample
A “Task Message” window will open. Press Yes to confirm sample
transfer.
Gate valve (V1) will open. Wait until white line has moved to “open”
position on gate valve to confirm before moving transfer rod.
Another “Task Message” window will open prompting to insert
sample. Carefully push transfer rod into chamber and sliding it
until the stop position. Fork should slide onto sample holder.
Push OK on Task Message Window.
3. Summitt -> Push Down [Z] in the “Intro” dialog box. The stage will move
Z down away from sample holder. The “Intro” dialog box will indicate that
the stage is moving.
If the stage is safely below the sample, carefully pull the transfer
rod out of the chamber to the stop position.
The gate valve (V1) will automatically close.
4. Watcher -> Press Backfill Intro. Will close valve over turbo pump and
open N2 gas to fill chamber.
With gloves on, remove sample and replace cap.
5. Watcher -> Press Pump Intro.
Ensure that Intro chamber is pumped before leaving tool.
Disable in Badger.
FIG 10: Summit “Image” Window, Stage Tab—Extract Procedure
(1)