1. Overview
This multimeter is characterized by a compact rugged construction with
protective holster and stand. The LCD Screen with 16mm display gives clear
readings even in low light when the back light can be switched on. Features
include sophisticated A/D converter in conjunction with overload protection.
Measures: DC & AC voltage, DC current Resistance and Temperature.
Tests: Continuity, Diode forward bias and Transistor hFE gain. Includes standard
probe test leads as well as temperature probe leads.
Before use you will need to fit a 9V PP3 battery (not supplied) in the compartment
located on the back of the meter. To open the compartment remove the cross
headed screw from the lid and lever open, clip the terminal studs over the battery
terminal caps, close the compartment lid and screw tightly shut.
2. Panel Layout
1 LCD Display: 3½ digits, character 16mm high
2 Data-hold Switch (HOLD).
3 Back Light Button Switch: Press this button to switch on back light in low light
conditions, the light will switch off automatically in 5 seconds. Press again to
switch it on again. If the battery power is low the light will dim.
4 Rotary Switch: use this switch to select functions and ranges
5 T+V Ω mA Input socket; 10A Input socket; T- COM Input socket
3. Safety Information
3-1 This meter is designed according to IEC-1010 for electronic measuring
instruments with an over-voltage category (CAT II) and pollution 2
3-2 Follow all safety and operating instructions to ensure that the meter is used
safely and is kept in good operating condition.
3-3 Safety symbols:
important safety information, refer to the operating manual.
Danger - high voltage.
Double insulation (protection Class II)
4. User Precautions
4-1 For safe use of the meter, use only the test leads supplied. If your test leads
are damaged make sure they are replaced with leads of the same quality and
specification.
4-2 Always use the meter with the blue holster in place.
4-3 Always check that the range switch is set to the correct position for the test
being made.
4-4 Do not exceed the maximum inputs shown on the dial or in these instructions.
Exceeding set inputs increases the chance of electric shock and can damage
your meter.
4-5 When measuring TV sets or switched power, avoid high voltage components
and high output pulse supplies.
4-6 During measurement be careful not to alter the Range switch position
accidentally.
4-7 Take extra care to avoid bodily contact with live surfaces when measuring
high voltage (especially more than DC 60V & AC 30V).
4-8 The Protection fuse should only be replaced with a fuse of the same type
and specification.
5. GENERAL SPECIFICATIONS
5-1 Max Voltage between input terminal and Earth Ground: CAT II 600V
5-2 Over-range Indication: display “1” for the significant digit.
5-3 Automatic display of negative polarity “_” .
5-4 Low Battery Indication: ‘
’ displayed
5-5 Max LCD display: 1999 (31/2 digits)
5-6 Fuse protection: F-200mA/250V (Ø5x20mm)
5-7 Power Supply: 9V battery, 6F22 or NEDA 1604
5-8 Operating Temp.: 0˚C to 40˚C (relative humidity <85%)
5-9 Storage Temp.:-1˚C to 50˚C (relative humidity <85%)
5-10 Guaranteed precision Temp.: 23±5 ˚C (relative humidity <85%)
5-11 Dimensions: 69x138x31mm (with holster)
5-12 Weight: approx. 170g (including battery)
6. Testing Specications
Accuracy is specified for a period of 1 year after calibration and a temperature
from 18˚C to 28˚C (64˚F to 82˚F) with relative humidity up to 75%.
6-1 DC Voltage
Range Resolution Accuracy
200mV 0.1mV ±(0.5% of rdg + 10 digit )
2V 1mV ±(0.5% of rdg + 10 digits)
20V 10mV ±(0.5% of rdg + 10 digits)
200V 100mV ±(0.8% of rdg + 10 digits)
600V 1V ±(1.2% of rdg + 10 digits)
- Input Impedance: 10MΩ
- Overload protection: 250V for 200mV range, effective DC or AC 600V for other
ranges
6-2 AC Voltage
Range Resolution accuracy
200V 100mV ±(1.2% of rdg + 20 digits)
600V 1V ±(1.2% of rdg + 20 digits)
- Frequency Range: 40 to 400Hz
- Response: average, calibrated in rms of sine wave
6-3 DC Current
Range Resolution Accuracy
2mA 1µA ±(1.0% of rdg + 10 digits)
20mA 10µA ±(1.0% of rdg + 10 digits)
200mA 100µA ±(1.5% of rdg + 20 digits)
10A 10mA ±(3.0% of rdg + 20 digits)
- Overload protection: F 200mA/250V fuse
Note: [1] 10A range: not fused
6- 4 Resistance
Range Resolution Accuracy
200Ω 0.1Ω ±(1.5% of rdg + 25 digits)
2KΩ ±(0.8% of rdg + 20 digits)
20KΩ 10Ω ±(0.8% of rdg + 20 digits)
200KΩ 100 ±(0.8% of rdg + 20 digits)
2MΩ 1KΩ ±(1.2% of rdg + 25 digits)
- Overload protection: 250V effective value
6-5 Temperature
Range Test Range Accuracy
-20~400 ˚C 1˚C ±(1.2% of rdg +5 digits)
400~1000 ˚C 1˚C ±(2.0% of rdg + 15 digits)
- Overload Protection: 250V DC or rms AC
6-6 Transistor hFE Test
Range Test Range Test Current / Voltage
NPN & PNP 0-1000 Ib=10uA / Vce=3V
6-7 Diode Test
Range Resolution Function
1mV
Display: read approximate forward
voltage of diode
- Overload Protection: 250V effective value
- Forward DC current: approximate 1mA
n Reversed DC voltage: approximate 3.0V
USER GUIDE
Philex Customer Careline 08457 573479
Digital Multimeter with
Temperature Measurement
83002R
Battery
Compartment
Insert 9V PP3
Question and Answer IconNeed help?

Do you have a question about the Philex 83002R and is the answer not in the manual?

Philex 83002R Specifications

General IconGeneral
Display3.5 digit LCD
DC Voltage Ranges200mV, 2V, 20V, 200V, 600V
AC Voltage Ranges200V, 600V
Resistance Ranges200Ω, 2kΩ, 20kΩ, 200kΩ, 2MΩ
Diode TestYes
Continuity TestYes
Battery Test1.5V, 9V
Overload ProtectionYes
Safety RatingCAT II 600V
Battery Type9V (6F22)
DC Current Ranges200μA, 2mA, 20mA, 200mA, 10A

Summary

Overview

Panel Layout

Safety Information

User Precautions

General Specifications

Testing Specifications

DC Voltage

Specifications for DC voltage measurement, accuracy, and overload protection.

AC Voltage

Specifications for AC voltage measurement, frequency range, and response.

DC Current

Specifications for DC current measurement, accuracy, and overload protection.

Resistance

Specifications for resistance measurement, accuracy, and overload protection.

Temperature

Specifications for temperature measurement range and overload protection.

Transistor hFE Test

Specifications for transistor hFE testing conditions and parameters.

Diode Test

Specifications for diode forward voltage drop measurement and polarity.

Operating Instructions

Attention before operation

Pre-operation checks, battery status indicators, and test lead safety.

Measuring DC Voltage

Step-by-step guide for measuring DC voltage, including range selection.

Measuring AC Voltage

Step-by-step guide for measuring AC voltage, including range selection.

Measuring DC Current

Step-by-step guide for measuring DC current, including lead connection.

Measuring Resistance

Step-by-step guide for measuring resistance, including circuit safety.

Measuring Temperature

Step-by-step guide for measuring temperature using the thermocouple.

Transistor Testing

Step-by-step guide for testing transistor hFE using the dedicated socket.

Diode Testing

Step-by-step guide for testing diodes, including polarity and readings.

Continuity Testing

Step-by-step guide for continuity testing, including buzzer indication.

Maintenance

Accessories