Complete Power-down (weekly):
1. A complete system power-down is comprised of three procedures:
iShut down the gantry and host computer
iShut down the CIRS server(s)
iShut down wall power
Recommendations for your system include:
iLeave the power on at all times; shut down weekly and restore power
after complete power-down is performed.
iShut down the gantry, host and CIRS once a week along with the
wall power.
Shutdown wall power:
Warning: Do not shut down wall power until the
gantry shuts down.
1. Locate the On/Off buttons (wall power).
2. Remove the power from the system by pressing the Off button.
The Universal Power Supply (UPS) will beep indicating power is off.
3. Wait at least 30 seconds before pressing On.
Note
Failure to wait at least 30 seconds may cause improper reboot, and
require
ou to repeat the entire process. Turn wall power back on after
30 seconds.
Note
For long-term shutdown (more than 10 minutes), turn off UPS.
Shutdown Gantry and Host computer:
1. Turn OFF the key on the scan control box. The gantry will usually
turn off in a few seconds; however, it can take up to 15 minutes if the
tube is at maximum heat (you do not need to wait for this shutdown
before shutting down the Host and CIRS). If you wish to abort the
shutdown, turn the key back to the ON position.
2. Click Logout. The system prompts you to confirm.
3. From the Home menu choose Logout. Confirm by clicking Yes.
4. At Windows Log In choose Shutdown, then Shutdown from the
options dropdown menu. Click OK. The system will shut down.
Note: To do a complete shutdown, proceed to shutdown CIR servers.
Tips
iIf you are unable to perform a function in any screen, log off the
system and log back on.
iIf you continue to have difficulties performing functions, reboot the
host computer.
Shut down the CIRS server(s):
1. Locate the CIRS server(s) inside the cabinet. Depending on your
system configuration, you will have up to three servers.
2. Press and hold the power button on each server (about 5 seconds)
until the green light is off. The bottom server may take longer to
shut down.