78
About Getting started Daily use Options Warnings More info
ITC/ITE HS/ITE FS
Measured according to American National Standard ANSI S3.22-2014,
IEC 60118-0:2015 and IEC 60318-5:2006
Supply voltage: Zinc-Air
Technical Data
0 dB SPL ref. 20 μPa HearLink 9030
HearLink 7030,
5030, 3030 & 2030
OSPL90
Peak 109 dB SPL 109 dB SPL
HF Average 103 dB SPL 103 dB SPL
Full-on Gain
Peak 46 dB 46 dB
HF Average 40 dB 40 dB
Reference Test Gain 26 dB 26 dB
Frequency Range 100-9400 Hz 100-7500 Hz
Total Harmonic
Distortion
500 Hz <2 % <2 %
800 Hz <2 % <2 %
1600 Hz <2 % <2 %
Equivalent Input
Noise Level
(omni/dir) 17/28 dB SPL 17/ 27 dB SPL
HF Average SPLITS (left/right ear) 85/85 dB SPL 85/85 dB SPL
Attack Time 5 ms 5 ms
Release Time 34 ms 33 ms