4 Configuration Settings Appendix Measurement Settings
34
Configuring ECG
Measurement Settings: Main Setup -> Measurements -> ECG
Factory Defaults
Item Name Oper.
Mode
MX800,
MP5 - MP90
(H10/20/40)
MP5T, MP5SC
MP2/X2
MX800,
MP5 - MP90 (H30)
MP2/X2
C M Profile
Adult
Profile
Pedi
Profile
Neo
Profile
Adult
Profile
Pedi
Profile
Neo
Profile
Outdoor
High Limit x x not applicable, see "Configuring ECG/Pulse Alarms" on page 38.
Low Limit x x
ECG/Arrhy Alarms x x
AlarmSrc (ECG/AR) x x
ECG x x On
Paced x x not applicable, the paced status is stored in a Profile, see "Profile Settings" on page 32
QRS Volume x x not applicable, this setting is stored in the Monitor Settings Block, see "Configuring
User Interface Settings" on page 126
Primary Lead x x II
Secondary Lead x x V (V1)
Va Lead x x V2
Vb Lead x x V5
Analysis Mode x x Multi Lead
Lead Placement x x Standard
Mod.LeadPlacment x x Off
Filter x x Monitor
Filter Monitor
Sync Out Chan 1 ECG AnlgWave (MX800 only); fixed setting; visible but not operable
Sync Out Chan 2 x ECG AnlgWave (MX800 only)
SyncPulse Sensit. x x Medium (MP2/X2, MP5 only)
SyncPulse Marker x x On (MP2/X2, MP5 only)
Auto Filter x not applicable, these settings are stored in the Monitor Settings Block, see "ECG
Application Configuration" on page 114
Fix PacerAmplit x
Default ECG Size x
Color x Green White
Asystole Threshold x x 4.0 sec 3.0 sec
4.0 sec
Δ ExtrTachy x not applicable, see "Configuring ECG/Pulse Alarms" on page 38
Tachy Clamp x
Δ ExtrBrady x
Brady Clamp x
ECG Al. OFF Inop x Cyan
Fallback x On